On the optimization of bone SPECT/CT in terms of image quality and radiation dose.
Tulik, Monika; Tulik, Piotr; Kowalska, Teresa.
Afiliação
  • Tulik M; Maria Sklodowska-Curie National Research Institute of Oncology Krakow Branch, Krakow, Poland.
  • Tulik P; Warsaw University of Technology, Faculty of Mechatronics, Institute of Metrology and Biomedical Engineering, Warsaw, Poland.
  • Kowalska T; Maria Sklodowska-Curie National Research Institute of Oncology Krakow Branch, Krakow, Poland.
J Appl Clin Med Phys ; 21(11): 237-246, 2020 Nov.
Article em En | MEDLINE | ID: mdl-33111500
ABSTRACT

INTRODUCTION:

The purpose of this study was to present the optimization process of CT parameters to reduce patient exposure during bone SPECT/CT without affecting the quality of SPECT images with attenuation correction (AC). MATERIAL AND

METHODS:

A fillable phantom reflecting realistic bone scintigraphy conditions was developed and acquired on an AnyScan SC. SPECT/CT scans were carried out with different x-ray tube current values (10, 20, 30, 40, 50, 60, 70, 90, 110, 130, 150, and 200 mA) at three different high-voltage values (80, 100, and 120 kV). The contrast (C) and coefficients of variation (CV) in the SPECT images as well as the signal-to-noise ratio (SNR) and noise (SDCT ) in the CT images with CTDIvol were measured. An optimal acquisition protocol that obtained SPECT/CT images with no artifacts on both CT and SPECT images, acceptable C, SNR, CV, and SDCT values, and the largest reduction in patient exposure compared to the reference acquisition procedure was sought.

RESULTS:

The optimal set of parameters for bone SPECT/CT was determined based on a phantom study. It has been implemented in clinical practice. Two groups of patients were examined according to the baseline and optimized protocols, respectively. The new SPECT/CT protocol substantially reduced patients' radiation exposure compared to the old protocol while maintaining the required diagnostic quality of SPECT and CT images.

CONCLUSIONS:

In the study, we present a methodology that finds a compromise between diagnostic information and patient exposure during bone SPECT/CT procedures.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Tomografia Computadorizada de Emissão de Fóton Único Tipo de estudo: Guideline Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article