Central venous pressure estimation from ultrasound assessment of the jugular venous pulse.
Zamboni, Paolo; Malagoni, Anna Maria; Menegatti, Erica; Ragazzi, Riccardo; Tavoni, Valentina; Tessari, Mirko; Beggs, Clive B.
Afiliação
  • Zamboni P; HUB Center Regione Emilia Romagna for Venous and Lymphatics Disorders, University Hospital of Ferrara, Ferrara, Italy.
  • Malagoni AM; HUB Center Regione Emilia Romagna for Venous and Lymphatics Disorders, University Hospital of Ferrara, Ferrara, Italy.
  • Menegatti E; HUB Center Regione Emilia Romagna for Venous and Lymphatics Disorders, University Hospital of Ferrara, Ferrara, Italy.
  • Ragazzi R; Department of Morphology, Surgery and Experimental Medicine, University of Ferrara, Ferrara, Italy.
  • Tavoni V; HUB Center Regione Emilia Romagna for Venous and Lymphatics Disorders, University Hospital of Ferrara, Ferrara, Italy.
  • Tessari M; HUB Center Regione Emilia Romagna for Venous and Lymphatics Disorders, University Hospital of Ferrara, Ferrara, Italy.
  • Beggs CB; Institute for Sport, Physical Activity and Leisure, Leeds Beckett University, Leeds, United Kingdom.
PLoS One ; 15(10): e0240057, 2020.
Article em En | MEDLINE | ID: mdl-33112871
ABSTRACT

OBJECTIVES:

Acquiring central venous pressure (CVP), an important clinical parameter, requires an invasive procedure, which poses risk to patients. The aim of the study was to develop a non-invasive methodology for determining mean-CVP from ultrasound assessment of the jugular venous pulse.

METHODS:

In thirty-four adult patients (age = 60 ± 12 years; 10 males), CVP was measured using a central venous catheter, with internal jugular vein (IJV) cross-sectional area (CSA) variation along the cardiac beat acquired using ultrasound. The resultant CVP and IJV-CSA signals were synchronized with electrocardiogram (ECG) signals acquired from the patients. Autocorrelation signals were derived from the IJV-CSA signals using algorithms in R (open-source statistical software). The correlation r-values for successive lag intervals were extracted and used to build a linear regression model in which mean-CVP was the response variable and the lagging autocorrelation r-values and mean IJV-CSA, were the predictor variables. The optimum model was identified using the minimum AIC value and validated using 10-fold cross-validation.

RESULTS:

While the CVP and IJV-CSA signals were poorly correlated (mean r = -0.018, SD = 0.357) due to the IJV-CSA signal lagging behind the CVP signal, their autocorrelation counterparts were highly positively correlated (mean r = 0.725, SD = 0.215). Using the lagging autocorrelation r-values as predictors, mean-CVP was predicted with reasonable accuracy (r2 = 0.612), with a mean-absolute-error of 1.455 cmH2O, which rose to 2.436 cmH2O when cross-validation was performed.

CONCLUSIONS:

Mean-CVP can be estimated non-invasively by using the lagged autocorrelation r-values of the IJV-CSA signal. This new methodology may have considerable potential as a clinical monitoring and diagnostic tool.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pressão Venosa Central / Veias Jugulares Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pressão Venosa Central / Veias Jugulares Tipo de estudo: Diagnostic_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2020 Tipo de documento: Article