Comparison of two real-time polymerase chain reaction assays for the detection of severe acute respiratory syndrome-CoV-2 from combined nasopharyngeal-throat swabs.
Siddiqui, Oves; Manchanda, Vikas; Yadav, Abhishek; Sagar, Tanu; Tuteja, Sanchita; Nagi, Nazia; Saxena, Sonal.
Afiliação
  • Siddiqui O; Department of Microbiology, Maulana Azad Medical College, New Delhi, India.
  • Manchanda V; Department of Microbiology, Maulana Azad Medical College, New Delhi, India.
  • Yadav A; Department of Microbiology, Maulana Azad Medical College, New Delhi, India.
  • Sagar T; Department of Microbiology, Maulana Azad Medical College, New Delhi, India.
  • Tuteja S; Department of Microbiology, Maulana Azad Medical College, New Delhi, India.
  • Nagi N; Department of Microbiology, Maulana Azad Medical College, New Delhi, India.
  • Saxena S; Department of Microbiology, Maulana Azad Medical College, New Delhi, India.
Indian J Med Microbiol ; 38(3 & 4): 385-389, 2020.
Article em En | MEDLINE | ID: mdl-33154251
ABSTRACT
CONTEXT In the absence of effective treatment or vaccine, the current strategy for the prevention of further transmission of severe acute respiratory syndrome (SARS) CoV-2 (COVID-19) infection is early diagnosis and isolation of cases. The diagnosis of SARS-CoV-2 is done by detecting viral RNA in the nasopharyngeal and throat swabs by real-time polymerase chain reaction (PCR). Many commercial assays are now available for performing the PCR assay.

AIMS:

The aim was to evaluate the performance of the SD Biosensor nCoV real-time detection kit with the real-time PCR kit provided by the Indian Council of Medical Research-National Institute of Virology (ICMR-NIV), Pune (NIV Protocol). SUBJECTS AND

METHODS:

A total of 253 pairs of nasopharyngeal-oropharyngeal swabs combined in a single viral transport medium were tested for viral RNA by both the protocols. The sensitivity and specificity of the SD Biosensor were calculated considering the ICMR-NIV kit as the gold standard. Matched pairs of recorded cycle threshold values (Ct values) were compared by Pearson's correlation coefficient.

RESULTS:

Concordant COVID-19 negative and positive PCR results were reported for 113 and 77 samples, respectively. The SD Biosensor kit additionally detected 62 cases, which were found negative by the NIV protocol. In all discordant positive results by the SD Biosensor kit, the average Ct values were higher than the concordant positive results. A total of forty samples tested positive for E gene by SD Biosensor and having Ct values <25 had 100% concordance with NIV protocol results and 39 samples tested positive for E gene by SD Biosensor having Ct value >32 were all found negative by the NIV protocol.

CONCLUSIONS:

The results highlight the need for careful evaluation of commercial kits before being deployed for screening of COVID-19 infections.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pneumonia Viral / Infecções por Coronavirus / Técnicas de Laboratório Clínico / Reação em Cadeia da Polimerase em Tempo Real / Betacoronavirus Tipo de estudo: Diagnostic_studies / Guideline / Screening_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Pneumonia Viral / Infecções por Coronavirus / Técnicas de Laboratório Clínico / Reação em Cadeia da Polimerase em Tempo Real / Betacoronavirus Tipo de estudo: Diagnostic_studies / Guideline / Screening_studies Limite: Humans Idioma: En Ano de publicação: 2020 Tipo de documento: Article