Automated digital enumeration of plasma cells in bone marrow trephine biopsies of multiple myeloma.
Malherbe, Jacques A J; Fuller, Kathryn A; Mirzai, Bob; Augustson, Bradley M; Erber, Wendy N.
Afiliação
  • Malherbe JAJ; School of Biomedical Sciences, Faculty of Health & Medical Sciences, The University of Western Australia, Crawley, Western Australia, Australia.
  • Fuller KA; School of Biomedical Sciences, Faculty of Health & Medical Sciences, The University of Western Australia, Crawley, Western Australia, Australia.
  • Mirzai B; School of Biomedical Sciences, Faculty of Health & Medical Sciences, The University of Western Australia, Crawley, Western Australia, Australia.
  • Augustson BM; PathWest Laboratory Medicine WA, Nedlands, Western Australia, Australia.
  • Erber WN; PathWest Laboratory Medicine WA, Nedlands, Western Australia, Australia.
J Clin Pathol ; 75(1): 50-57, 2022 Jan.
Article em En | MEDLINE | ID: mdl-33234694
ABSTRACT

AIMS:

Determination of the number of plasma cells in bone marrow biopsies is required for the diagnosis and ongoing evaluation of plasma cell neoplasms. We developed an automated digital enumeration platform to assess plasma cells identified by antigen expression in whole bone marrow sections in multiple myeloma, and compared it with manual assessments.

METHODS:

Bone marrow trephine biopsy specimens from 91 patients with multiple myeloma at diagnosis, remission and relapse were stained for CD138 and multiple myeloma oncogene 1 (MUM1). Manual assessment and digital quantification were performed for plasma cells in the entire trephine section. Concordance rates between manual and digital methods were evaluated for each antigen by intraclass correlation analyses (ICC) with associated Spearman's correlations.

RESULTS:

The digital platform counted 16 484-1 118 868 cells and the per cent CD138 and MUM1-positive plasma cells ranged from 0.05% to 93.5%. Overall concordance between digital and manual methods was 0.63 for CD138 and 0.89 for MUM1. Concordance was highest with diffuse plasma cell infiltrates (MUM1 ICC=0.90) and lowest when in microaggregates (CD138 ICC=0.13). Manual counts exceeded digital quantifications for both antigens (CD138 mean=26.4%; MUM1 mean=9.7%). Diagnostic or relapse threshold counts, as determined by CD138 manual assessments, were not reached with digital counting for 16 cases (18%).

CONCLUSIONS:

Automated digital enumeration of the entire, immunohistochemically stained bone marrow biopsy section can accurately determine plasma cell burden, irrespective of pattern and extent of disease (as low as 0.05%). This increases precision over manual visual assessments which tend to overestimate plasma burden, especially for CD138, and when plasma cells are in clusters.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Mieloma Múltiplo Tipo de estudo: Guideline / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Mieloma Múltiplo Tipo de estudo: Guideline / Prognostic_studies Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article