Optimization of Analytical Conditions for Hydrophilic Nucleic Acids Using Mixed-Mode and Reversed-Phase Pentabromobenzyl Columns.
Hosokawa, Mika; Goto, Kanako; Tanaka, Shota; Ueda, Kumiko; Iwakawa, Seigo; Ogawara, Ken-Ichi.
Afiliação
  • Hosokawa M; Department of Pharmaceutics, Kobe Pharmaceutical University.
  • Goto K; Department of Pharmaceutics, Kobe Pharmaceutical University.
  • Tanaka S; Department of Pharmaceutics, Kobe Pharmaceutical University.
  • Ueda K; Department of Pharmaceutics, Kobe Pharmaceutical University.
  • Iwakawa S; Department of Pharmaceutics, Kobe Pharmaceutical University.
  • Ogawara KI; Department of Pharmaceutics, Kobe Pharmaceutical University.
Chem Pharm Bull (Tokyo) ; 68(12): 1233-1237, 2020.
Article em En | MEDLINE | ID: mdl-33268655
The aim of this study was to investigate appropriate analytical conditions for hydrophilic nucleosides and nucleotides (monophosphates and triphosphates) by HPLC methods using a mixed-mode AX-C18 column with anion-exchange and hydrophobic interactions by quaternary ammonium and C18, respectively, and a reversed-phase pentabromobenzyl (PBr) column with dispersion force and hydrophobic interactions by PBr group. The higher compound polarity led to stronger retention on AX-C18 (triphosphates > monophosphates > nucleosides). AX-C18 demonstrated feasible retention of nucleotides via anion-exchange interaction by increasing the salt and methanol concentrations. In contrast, on PBr, the lower compound polarity led to stronger retention. On PBr, feasible retention of both nucleosides and nucleotides was obtained via dispersion interactions with purine and pyrimidine rings by increasing the methanol concentration. Regarding the pH of phosphate buffer used as the mobile phase, pH 7.0 should be used in measuring nucleoside triphosphates on AX-C18, whereas pH 2.5 is better suited for measuring nucleotides on PBr. In terms of selectivity to highly hydrophilic nucleotides, the mixed-mode AX-C18 column had an advantage over the reverse-phase PBr column. In contrast, PBr column was more versatile than the AX-C18 column. Taken together, HPLC analyses of nucleosides and nucleotides should be carried out by optimizing the interactions between the stationary phase and nucleic acids.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fosfatos / Ácidos Nucleicos Idioma: En Ano de publicação: 2020 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fosfatos / Ácidos Nucleicos Idioma: En Ano de publicação: 2020 Tipo de documento: Article