Your browser doesn't support javascript.
loading
Real-World Experience of the MANTA Closure Device: Insights From the FDA Manufacturer and User Facility Device Experience (MAUDE) Database.
Case, Brian C; Yerasi, Charan; Forrestal, Brian J; Kumar, Sant; Musallam, Anees; Chezar-Azerrad, Chava; Khalid, Nauman; Shlofmitz, Evan; Khan, Jaffar M; Satler, Lowell F; Ben-Dor, Itsik; Rogers, Toby; Waksman, Ron.
Afiliação
  • Case BC; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America.
  • Yerasi C; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America.
  • Forrestal BJ; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America.
  • Kumar S; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America.
  • Musallam A; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America.
  • Chezar-Azerrad C; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America.
  • Khalid N; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America.
  • Shlofmitz E; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America.
  • Khan JM; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America; Cardiovascular Branch, Division of Intramural Research, National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, MD, United States of America.
  • Satler LF; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America.
  • Ben-Dor I; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America.
  • Rogers T; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America; Cardiovascular Branch, Division of Intramural Research, National Heart, Lung and Blood Institute, National Institutes of Health, Bethesda, MD, United States of America.
  • Waksman R; Section of Interventional Cardiology, MedStar Washington Hospital Center, Washington, DC, United States of America. Electronic address: ron.waksman@medstar.net.
Cardiovasc Revasc Med ; 27: 63-66, 2021 06.
Article em En | MEDLINE | ID: mdl-33402323
BACKGROUND/PURPOSE: The MANTA vascular closure device (VCD) is the first commercially available dedicated closure device for large-bore femoral arterial access-site closure and was approved by the United States Food and Drug Administration (FDA) in February 2019. Real-world data on the most commonly reported complications and modes of failure associated with the MANTA closure device are limited. We analyzed post-marketing surveillance data from FDA's Manufacturer and User Facility Device Experience (MAUDE) database for the MANTA VCD (Teleflex, Wayne, Pennsylvania). METHODS/MATERIALS: Post-marketing surveillance data from the FDA MAUDE database from February 2019 through March 2020 were analyzed, yielding 170 reports. RESULTS: Of the 170 reports of major complications involving MANTA devices, 141 reports involved either injury (136) or death (5) related to the device. Of the 141 reports, bleeding was the most common adverse outcome described (45), followed by vessel occlusion (30) and vessel dissection (21). Device malfunction incidents (29 reports) were reported in 4 main categories: failed deployment (16 reports), malposition of the collagen (9), insufficient information (3), and device dislocation (1). CONCLUSIONS: Our analysis of the MAUDE database demonstrates that in real-world practice, the MANTA VCD was found to be associated with complications, including death, vascular injury, and difficulties with the device itself. Ongoing user education, proctoring, and pre-procedural patient selection are important to minimize risks associated with the MANTA VCD.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dispositivos de Oclusão Vascular Tipo de estudo: Etiology_studies Limite: Humans País como assunto: America do norte Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Dispositivos de Oclusão Vascular Tipo de estudo: Etiology_studies Limite: Humans País como assunto: America do norte Idioma: En Ano de publicação: 2021 Tipo de documento: Article