How accurately can surgeons perform angle manipulation? Quantitative assessment of the accuracy of manual angle manipulation of orthopedic surgery: a cadaver study.
Park, Young Hwan; Kim, Woon; Choi, Jung Woo; Choi, Gi Won; Kim, Hak Jun.
Afiliação
  • Park YH; Department of Orthopedic Surgery, Guro Hospital, Korea University, 148 Gurodong-ro, Guro-gu, Seoul, 08308, Korea.
  • Kim W; Department of Orthopedic Surgery, Guro Hospital, Korea University, 148 Gurodong-ro, Guro-gu, Seoul, 08308, Korea.
  • Choi JW; Department of Orthopedic Surgery, Guro Hospital, Korea University, 148 Gurodong-ro, Guro-gu, Seoul, 08308, Korea.
  • Choi GW; Department of Orthopedic Surgery, Ansan Hospital, Korea University, 123 Jeokgeum-ro, Danwon-gu, Ansan, 15355, Korea.
  • Kim HJ; Department of Orthopedic Surgery, Guro Hospital, Korea University, 148 Gurodong-ro, Guro-gu, Seoul, 08308, Korea. hjunkimos@gmail.com.
Arch Orthop Trauma Surg ; 142(6): 905-911, 2022 Jun.
Article em En | MEDLINE | ID: mdl-33417029
ABSTRACT

INTRODUCTION:

To date, only few studies have been performed on the accuracy of manual angle manipulation during orthopedic surgery. This cadaver study was aimed at quantitatively assessing the accuracy of manual angle manipulation performed by orthopedic surgeons according to their surgical experience and comparing it with manipulation performed with the assistance of a digital goniometer. MATERIALS AND

METHODS:

Six lower-leg specimens of fresh-frozen human cadavers were subjected to angle manipulation performed via Kirschner wire (K-wire) insertion. K-wires were inserted manually and with the assistance of a digital goniometer at target angles of 0°, 30°, and 60° by three operators who had different levels of experience in orthopedic surgery. The accuracy of the insertion angles at the target angles was evaluated using computed tomography.

RESULTS:

The mean angle error in the manual angle manipulation was 8.8° (standard deviation [SD] 6.0). When the target angles were set to 0°, 30°, and 60°, the identified angle errors were 6.1° (SD 4.3), 8.8° (SD 6.6), and 11.7° (SD 5.6), respectively, and each value did not show any significant difference among the operators. With the assistance of a digital goniometer, the mean (SD) angle error was significantly improved to 2.1° (1.1°) (p < 0.001). The amount of improvement in accuracy significantly increased as the target angle increased (p = 0.01).

CONCLUSION:

This cadaver study quantified the inaccuracy of manual angle manipulation in orthopedic surgery and showed that these inaccuracies ​​can be improved using an assistive device. These results support the need to develop a device that can compensate manual angle manipulation in orthopedic surgery.