Beat-to-beat blood pressure measurement using a cuffless device does not accurately reflect invasive blood pressure.
Moharram, Mohammed A; Wilson, Luke C; Williams, Michael Ja; Coffey, Sean.
Afiliação
  • Moharram MA; Department of Medicine - HeartOtago, Dunedin School of Medicine, University of Otago, Dunedin, New Zealand.
  • Wilson LC; Department of Medicine - HeartOtago, Dunedin School of Medicine, University of Otago, Dunedin, New Zealand.
  • Williams MJ; Department of Medicine - HeartOtago, Dunedin School of Medicine, University of Otago, Dunedin, New Zealand.
  • Coffey S; Department of Medicine - HeartOtago, Dunedin School of Medicine, University of Otago, Dunedin, New Zealand.
Int J Cardiol Hypertens ; 5: 100030, 2020 Jun.
Article em En | MEDLINE | ID: mdl-33447759
ABSTRACT

BACKGROUND:

The availability of an accurate continuous cuffless blood pressure (BP) monitor would provide an alternative to both invasive continuous BP and 24-h intermittent cuff-based BP monitors. We investigated the accuracy of a cuffless beat to beat (BtB) device compared to both invasive BP (iBP) and brachial cuff BP (cBP) measurements.

METHODS:

Patients undergoing clinically indicated coronary angiography (CA) and/or percutaneous coronary intervention (PCI) were recruited. After calibration to an initial cBP reading, BP was measured simultaneously using a BtB device (SOMNOtouch NIBP), brachial artery iBP, and cBP at two time points.

RESULTS:

The study was terminated early due to a significant bias. Recordings from 14 participants (11 males, mean age 68.4 years) were analysed. Readings from BtB BP were higher than iBP. The bias between BtB BP and iBP was 34.3 mmHg (95%CI 27.0, 41.5) and 23.6 mmHg (95%CI 16.8, 30.4) for SBP and DBP respectively. A similar bias was seen between BtB BP and cBP, but cBP and iBP were largely in agreement.

CONCLUSIONS:

In patients undergoing CA/PCI, significant differences were detected between BtB BP and both invasively measured and cuff BP. The non-invasive BtB BP measurement device tested is not suitable for clinical or research use.