Wire mesh capped DRPS based bronchial dosimeter for personal inhalation dosimetry due to radon progeny.
J Radiol Prot
; 41(2)2021 06 01.
Article
em En
| MEDLINE
| ID: mdl-33571976
A study has been carried out to experimentally determine the calibration factor (CF) of the passive bronchial dosimeter, which consists of a direct radon progeny sensor capped with a 100-wire mesh. First, the CF was determined in controlled environmental conditions simulated in a calibration chamber. With aerosol concentrations varying from 104p cm-3to 105p cm-3and relative humidity varying from 60% to 80% in the chamber, CF was observed to be nearly constant with an average value of (3.8 ± 0.5) × 10-3mSv tracks-1cm2. Then, the CF was determined in real indoor environments in which it was again observed to be almost constant and the mean value was found to be (5.6 ± 0.1) × 10-3mSv tracks-1cm2. Pooling all the data on CFs obtained under controlled conditions and in real indoor environments, a lognormal distribution of the CF was observed with a geometric mean and geometric standard deviation of 0.0052 mSv tracks-1cm2and 1.28 respectively. The experimentally determined value of CF was found to be in close agreement with the theoretically estimated value, taking into consideration the unattached fraction of radon progeny. This dosimeter is passive, cheap, lightweight and, moreover, the CF being stable against environmental variations, will be useful in monitoring inhalation doses due to radon progeny for occupational workers.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Monitoramento de Radiação
/
Radônio
/
Poluição do Ar em Ambientes Fechados
/
Poluentes Radioativos do Ar
Limite:
Humans
Idioma:
En
Ano de publicação:
2021
Tipo de documento:
Article