Testing the Limits of a Portable NIR Spectrometer: Content Uniformity of Complex Powder Mixtures Followed by Calibration Transfer for In-Line Blend Monitoring.
Casian, Tibor; Gavan, Alexandru; Iurian, Sonia; Porfire, Alina; Toma, Valentin; Stiufiuc, Rares; Tomuta, Ioan.
Afiliação
  • Casian T; Department of Pharmaceutical Technology and Biopharmacy, Faculty of Pharmacy, "Iuliu Hatieganu" University of Medicine and Pharmacy, 400012 Cluj-Napoca, Romania.
  • Gavan A; Department of Medical Devices, Faculty of Pharmacy, "Iuliu Hatieganu" University of Medicine and Pharmacy, 400349 Cluj-Napoca, Romania.
  • Iurian S; Department of Pharmaceutical Technology and Biopharmacy, Faculty of Pharmacy, "Iuliu Hatieganu" University of Medicine and Pharmacy, 400012 Cluj-Napoca, Romania.
  • Porfire A; Department of Pharmaceutical Technology and Biopharmacy, Faculty of Pharmacy, "Iuliu Hatieganu" University of Medicine and Pharmacy, 400012 Cluj-Napoca, Romania.
  • Toma V; MedFuture Research Center for Advanced Medicine, "Iuliu Hațieganu" University of Medicine and Pharmacy, 400337 Cluj-Napoca, Romania.
  • Stiufiuc R; MedFuture Research Center for Advanced Medicine, "Iuliu Hațieganu" University of Medicine and Pharmacy, 400337 Cluj-Napoca, Romania.
  • Tomuta I; Department of Pharmaceutical Technology and Biopharmacy, Faculty of Pharmacy, "Iuliu Hatieganu" University of Medicine and Pharmacy, 400012 Cluj-Napoca, Romania.
Molecules ; 26(4)2021 Feb 20.
Article em En | MEDLINE | ID: mdl-33672675
ABSTRACT
(1)

Background:

Portable NIR spectrometers gain more and more ground in the field of Process Analytical Technology due to the easy on-site flexibility and interfacing versatility. These advantages that originate from the instrument miniaturization, also come with a downside with respect to performance compared to benchtop devices. The objective of this work was to evaluate the performance of MicroNIR in a pharmaceutical powder blend application, having three active ingredients and 5 excipients. (2)

Methods:

Spectral data was recorded in reflectance mode using static and dynamic acquisition, on calibration set samples developed using an experimental design. (3)

Results:

The developed method accurately predicted the content uniformity of these complex mixtures, moreover it was validated in the entire calibration range using ±10% acceptance limits. With respect to at-line prediction, the method presented lower performance compared to a previously studied benchtop spectrometer. Regarding the in-line monitoring of the blending process, it was shown that the spectral variability-induced by dynamic acquisition could be efficiently managed using spectral pre-processing. (4)

Conclusions:

The in-line process monitoring resulted in accurate concentration profiles, highlighting differences in the mixing behaviour of the investigated ingredients. For the low dose component homogeneity was not reached due to an inefficient dispersive mixing.