Establishment of rapid, sensitive, and quantitative liquid chromatography-electrospray ionization-tandem mass spectrometry method coupled with liquid-liquid extraction for measurement of urinary bisphenol A, 4-t-octylphenol, and 4-nonylphenol.
Battal, Dilek; Sukuroglu, Ayca Aktas; Kocadal, Kumsal; Cok, Ismet; Unlusayin, Irfan.
Afiliação
  • Battal D; Department of Toxicology, Faculty of Pharmacy, Mersin University, Mersin, Turkey.
  • Sukuroglu AA; Department of Toxicology, Faculty of Pharmacy, Near East University, Nicosia, Cyprus.
  • Kocadal K; Department of Toxicology, Faculty of Pharmacy, Gazi University, Ankara, Turkey.
  • Cok I; Department of Toxicology, Faculty of Pharmacy, Near East University, Nicosia, Cyprus.
  • Unlusayin I; Department of Toxicology, Faculty of Pharmacy, Gazi University, Ankara, Turkey.
Rapid Commun Mass Spectrom ; 35(12): e9084, 2021 Jun 30.
Article em En | MEDLINE | ID: mdl-33733529
ABSTRACT
RATIONALE Bisphenol A (BPA), 4-t-octylphenol (4-t-OP), and 4-nonylphenol (4-NP) are ubiquitous manufactured chemicals classified as endocrine disruptors. Now these substances are prevalent and found in a wide range of biological and environmental matrices globally. Acting as xenoestrogens, their target is to bind estrogen receptors and compete against natural hormones. Thus, there is an increasing concern on their negative health consequences.

METHODS:

A rapid, sensitive, and robust method was established, validated, and used for the trace analysis of BPA, 4-t-OP, and 4-NP in human urine using liquid chromatography-electrospray ionization-tandem mass spectrometry.

RESULTS:

The linear ranges for BPA, 4-t-OP, and 4-NP were 5.0-500 ng/mL. The correlation coefficient was higher than 0.996. The limit of detection and limit of quantitation for the compounds were 0.001, 0.007, and 0.005 ng/mL and 0.005, 0.023, and 0.180 ng/mL, respectively. The recoveries were as follows 99.539%-108.557%, 97.780%-110.768%, and 99.694%-106.908% for BPA, 4-NP, and 4-t-OP respectively, with relative standard deviation <8.26%.

CONCLUSIONS:

The established method was also applied effectively for the determination of BPA, 4-t-OP, and 4-NP levels in 15 human urine samples. This method was specifically developed for easy and precise analysis of BPA, 4-t-OP, and 4-NP in human urine. The method developed has been shown to be accurate, precise, and sensitive.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fenóis / Compostos Benzidrílicos / Cromatografia Líquida / Espectrometria de Massas em Tandem / Extração Líquido-Líquido Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fenóis / Compostos Benzidrílicos / Cromatografia Líquida / Espectrometria de Massas em Tandem / Extração Líquido-Líquido Tipo de estudo: Diagnostic_studies / Evaluation_studies Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article