Free-breathing simultaneous T1 , T2 , and T2 quantification in the myocardium.
Hermann, Ingo; Kellman, Peter; Demirel, Omer B; Akçakaya, Mehmet; Schad, Lothar R; Weingärtner, Sebastian.
Afiliação
  • Hermann I; Department of Imaging Physics, Magnetic Resonance Systems Lab, Delft University of Technology, Delft, The Netherlands.
  • Kellman P; Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany.
  • Demirel OB; National Heart, Lung, and Blood Institute, National Institutes of Health, DHHS, Bethesda, MD, USA.
  • Akçakaya M; Department of Electrical and Computer Engineering and Center for Magnetic Resonance Research, University of Minnesota, Minnesota, MN, USA.
  • Schad LR; Department of Electrical and Computer Engineering and Center for Magnetic Resonance Research, University of Minnesota, Minnesota, MN, USA.
  • Weingärtner S; Computer Assisted Clinical Medicine, Medical Faculty Mannheim, Heidelberg University, Mannheim, Germany.
Magn Reson Med ; 86(3): 1226-1240, 2021 09.
Article em En | MEDLINE | ID: mdl-33780037
ABSTRACT

PURPOSE:

To implement a free-breathing sequence for simultaneous quantification of T1 , T2 , and T2∗ for comprehensive tissue characterization of the myocardium in a single scan using a multi-gradient-echo readout with saturation and T2 preparation pulses.

METHODS:

In the proposed Saturation And T2 -prepared Relaxometry with Navigator-gating (SATURN) technique, a series of multi-gradient-echo (GRE) images with different magnetization preparations was acquired during free breathing. A total of 35 images were acquired in 26.5 ± 14.9 seconds using multiple saturation times and T2 preparation durations and with imaging at 5 echo times. Bloch simulations and phantom experiments were used to validate a 5-parameter fit model for accurate relaxometry. Free-breathing simultaneous T1 , T2 , and T2∗ measurements were performed in 10 healthy volunteers and 2 patients using SATURN at 3T and quantitatively compared to conventional single-parameter methods such as SASHA for T1 , T2 -prepared bSSFP, and multi-GRE for T2∗ .

RESULTS:

Simulations confirmed accurate fitting with the 5-parameter model. Phantom measurements showed good agreement with the reference methods in the relevant range for in vivo measurements. Compared to single-parameter methods comparable accuracy was achieved. SATURN produced in vivo parameter maps that were visually comparable to single-parameter methods. No significant difference between T1 , T2 , and T2∗ times acquired with SATURN and single-parameter methods was shown in quantitative measurements (SATURN T1=1573±86ms , T2=33.2±3.6ms , T2∗=25.3±6.1ms ; conventional

methods:

T1=1544±107ms , T2=33.2±3.6ms , T2∗=23.8±5.5ms ; P>.2 )

CONCLUSION:

SATURN enables simultaneous quantification of T1 , T2 , and T2∗ in the myocardium for comprehensive tissue characterization with co-registered maps, in a single scan with good agreement to single-parameter methods.