Ultrasound-Guided Deployment of a Plug-Based Vascular Closure Device in Femoral Arteries with Calcified Plaque or Stenosis.
J Vasc Interv Radiol
; 32(6): 802-806, 2021 06.
Article
em En
| MEDLINE
| ID: mdl-34051988
From January 2019 to January 2020, 106 patients (age, 64.8 ± 14.1 years; male, 63.2%) were included to retrospectively investigate the feasibility and safety of ultrasound-guided deployment of ExoSeal after femoral artery access. Baseline characteristics were not different except for age (P = .022), body mass index (P = .009), and diameter (P < .001) between the calcified plaque or stenosis (CS) group (n = 49) and non-CS group (n = 57). The overall technical and clinical success rates were 96.2% and 100%, respectively. The technical (CS group, 48/49; non-CS group, 54/57) and clinical success rates (100%), time to hemostasis (CS group, 3.21 ± 0.54 min; non-CS group, 3.39 ± 0.71 min), and complication rates (CS group, 1/49; non-CS group, 0/57) were not different between the 2 groups. ExoSeal seems to be safe to use under ultrasound guidance in the femoral arteries with CS.
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Cateterismo Periférico
/
Técnicas Hemostáticas
/
Ultrassonografia de Intervenção
/
Artéria Femoral
/
Doença Arterial Periférica
/
Calcificação Vascular
/
Dispositivos de Oclusão Vascular
Tipo de estudo:
Guideline
/
Observational_studies
Limite:
Aged
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Ano de publicação:
2021
Tipo de documento:
Article