Your browser doesn't support javascript.
loading
Utility of process capability indices in assessment of quality control processes at a clinical laboratory chain.
Dong, Ping; Wang, Yong-Bo; Peng, De-Zhi; Wang, Jia-Jia; Cheng, Ya-Ting; Deng, Xiao-Yan; Zheng, Biao; Tao, Ran.
Afiliação
  • Dong P; Laboratory Diagnosis Department, Jinan KingMed Center for Clinical Laboratory, Jinan, China.
  • Wang YB; Laboratory Diagnosis Department, Qingdao KingMed Center for Clinical Laboratory, Qingdao, China.
  • Peng DZ; Laboratory Diagnosis Department, Jinan KingMed Center for Clinical Laboratory, Jinan, China.
  • Wang JJ; Laboratory Diagnosis Department, Jinan KingMed Center for Clinical Laboratory, Jinan, China.
  • Cheng YT; Laboratory Diagnosis Department, Guangzhou KingMed Center for Clinical Laboratory, Guangzhou, China.
  • Deng XY; KingMed School of Laboratory Medicine, Guangzhou Medical University, Guangzhou, China.
  • Zheng B; KingMed School of Laboratory Medicine, Guangzhou Medical University, Guangzhou, China.
  • Tao R; KingMed School of Laboratory Medicine, Guangzhou Medical University, Guangzhou, China.
J Clin Lab Anal ; 35(8): e23878, 2021 Aug.
Article em En | MEDLINE | ID: mdl-34165837
ABSTRACT

BACKGROUND:

To evaluate the utility of the process capability indices Cp and Cpk for assessing the quality control processes at chain laboratory facilities.

METHODS:

In April 2020, the minimum Cp and Cpk values for 33 assays of a laboratory chain with 19 facilities were collected for further analysis and a total of 627 datasets (Cp and Cpk ) were compared. In addition, standard values for Cp and Cpk , defined as the lowest of the top 20%, were obtained for comparison and the indices were used to determine whether precision or trueness improvements were required for the corresponding assay.

RESULTS:

A total of 627 datasets of 33 assays from 19 laboratory facilities were collected for further analysis. Based on the Cp results, 329 (52.5%), 211 (33.7%), 65 (10.3%), and 22 (3.5%) were rated as excellent, good, marginal, and poor, respectively. While the corresponding results for Cpk were 300 (47.8%), 216 (34.4%), 79 (12.6%), and 32 (5.1%). In addition, it was noteworthy that eight (Cp criteria) and six assays (Cpk criteria) were rated as excellent or good at all 19 facilities. Comparison of the process capability indices at the Jinan KingMed Center with the standard values revealed that total protein, albumin, and urea showed trueness individual improvement, precision individual improvement, and precision common improvement, respectively, while the results of other assays were stable.

CONCLUSION:

Process capability indices are useful for evaluating the quality control procedures in laboratory facilities and can help improve the precision and trueness of laboratory tests.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Controle de Qualidade / Laboratórios Clínicos Tipo de estudo: Prognostic_studies Limite: Humans País como assunto: Asia Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Controle de Qualidade / Laboratórios Clínicos Tipo de estudo: Prognostic_studies Limite: Humans País como assunto: Asia Idioma: En Ano de publicação: 2021 Tipo de documento: Article