Feasibility and Significance of Dose Adaptation via Linear Couch Translations to Correct for Rotational Shifts During Frameless Brain Radiosurgery with the Gamma Knife Icon™.
Cheung, Joey P; Morin, Olivier; Braunstein, Steve E; Sneed, Penny K; Theodosopoulos, Philip V; McDermott, Michael W; Ma, Lijun.
Afiliação
  • Cheung JP; Department of Radiation Oncology, University of California San Francisco (UCSF), San Francisco, CA, USA.
  • Morin O; Department of Radiation Oncology, University of California San Francisco (UCSF), San Francisco, CA, USA.
  • Braunstein SE; Department of Radiation Oncology, University of California San Francisco (UCSF), San Francisco, CA, USA.
  • Sneed PK; Department of Radiation Oncology, University of California San Francisco (UCSF), San Francisco, CA, USA.
  • Theodosopoulos PV; Department of Neurosurgery, University of California San Francisco (UCSF), San Francisco, CA, USA.
  • McDermott MW; Department of Neurosurgery, University of California San Francisco (UCSF), San Francisco, CA, USA.
  • Ma L; Department of Radiation Oncology, University of California San Francisco (UCSF), San Francisco, CA, USA. lijun.ma@ucsf.edu.
Acta Neurochir Suppl ; 128: 145-150, 2021.
Article em En | MEDLINE | ID: mdl-34191071
OBJECTIVE: The present study aimed to examine the technical feasibility and effectiveness of adapting the radiation dose distributions with three-dimensional (3D) linear couch translations in contrast to full six-dimensional couch maneuvers to correct for rotational shifts during frameless radiosurgical treatment with the Gamma Knife Icon™ (Elekta AB; Stockholm, Sweden). METHODS: The original magnetic resonance images used for radiosurgery treatment planning (15 targets) were digitally processed to simulate rotational shifts of ±1, ±2, ±3, ±5, and ±10 degrees in the transverse plane and imported back into Leksell GammaPlan® (Elekta AB), creating "uncorrected" treatment plans. In addition, geometrically optimized 3D translation shifts were consequently applied to each isocenter in all "uncorrected" treatment plans to account for systematically introduced rotational shifts and to produce "corrected" treatment plans. The differences in the dose distribution between the original treatment plans and the "uncorrected" and "corrected" treatment plans were calculated and compared at each rotational shift position. RESULTS: The "uncorrected" treatment plans resulted in a significant deterioration in target coverage (by 8-72%) and selectivity (by 2-42%), with some targets being missed completely with rotations of ±3 or more degrees. In contrast, in all "corrected" treatment plans, the average decreases in target coverage and selectivity were only 1% (maximum values 4-5%). CONCLUSION: Applications of 3D linear couch translations successfully overcome gross uncertainties in dose distributions caused by up to ±10 degrees of rotational shifts in a target. As a result, rapid dose adaptation with 3D couch translations is unique and effective for frameless radiosurgery with the Gamma Knife Icon™.