Your browser doesn't support javascript.
loading
Quantitation of Cabozantinib in Human Plasma by LC-MS/MS.
Jones, Reyna; Holleran, Julianne; Parise, Robert A; Rudek, Michelle A; Chan, Jennifer; Wen, Yujia; Gobburu, Joga; Lewis, Lionel D; Beumer, Jan H.
Afiliação
  • Jones R; Cancer Therapeutics Program, UPMC Hillman Cancer Center, 5115 Centre Ave, Pittsburgh, PA 15232 , USA.
  • Holleran J; Cancer Therapeutics Program, UPMC Hillman Cancer Center, 5115 Centre Ave, Pittsburgh, PA 15232 , USA.
  • Parise RA; Cancer Therapeutics Program, UPMC Hillman Cancer Center, 5115 Centre Ave, Pittsburgh, PA 15232 , USA.
  • Rudek MA; Department of Oncology and Medicine, Johns Hopkins University School of Medicine and Johns Hopkins Sidney Kimmel Cancer Center, 1800 Orleans St, Baltimore, MD 21287, USA.
  • Chan J; Dana Farber/Partners CancerCare, Dana-Farber Cancer Institute 450 Brookline Ave. Boston, MA 02215-5450, USA.
  • Wen Y; Alliance for Clinical Trials in Oncology, 125 S. Wacker Drive, Suite 1600, Chicago, IL 60606 , USA.
  • Gobburu J; Center for Translational Medicine, University of Maryland, 20 North Pine Street Baltimore, Maryland 21201 , USA.
  • Lewis LD; Norris Cotton Cancer Center, Dartmouth-Hitchcock Medical Center, The Geisel School of Medicine at Dartmouth, 1 Medical Center Dr, Lebanon, NH 03766, USA.
  • Beumer JH; Cancer Therapeutics Program, UPMC Hillman Cancer Center, 5115 Centre Ave, Pittsburgh, PA 15232 , USA.
J Chromatogr Sci ; 60(3): 274-279, 2022 Mar 23.
Article em En | MEDLINE | ID: mdl-34240176
ABSTRACT
To support a phase III randomized trial of the multi-targeted tyrosine kinase inhibitor cabozantinib in neuroendocrine tumors, we developed a high-performance liquid chromatography mass spectrometry method to quantitate cabozantinib in 50 µL of human plasma. After acetonitrile protein precipitation, chromatographic separation was achieved with a Phenomenex synergy polar reverse phase (4 µm, 2 × 50 mm) column and a gradient of 0.1% formic acid in acetonitrile and 0.1% formic acid in water over a 5-min run time. Detection was performed on a Quattromicro quadrupole mass spectrometer with electrospray, positive-mode ionization. The assay was linear over the concentration range 50-5000 ng/mL and proved to be accurate (103.4-105.4%) and precise (<5.0%CV). Hemolysis (10% RBC) and use of heparin as anticoagulant did not impact quantitation. Recovery from plasma varied between 103.0-107.7% and matrix effect was -47.5 to -41.3%. Plasma freeze-thaw stability (97.7-104.9%), stability for 3 months at -80°C (103.4-111.4%), and stability for 4 h at room temperature (100.1-104.9%) were all acceptable. Incurred sample reanalysis of (N = 64) passed 100% samples within 20% difference, -0.7% median difference and 1.1% median absolute difference. External validation showed a bias of less than 1.1%. This assay will help further define the clinical pharmacokinetics of cabozantinib.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Espectrometria de Massas em Tandem / Anilidas Tipo de estudo: Clinical_trials Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Espectrometria de Massas em Tandem / Anilidas Tipo de estudo: Clinical_trials Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article