Your browser doesn't support javascript.
loading
Robust arterial input function surrogate measurement from the superior sagittal sinus complex signal for fast dynamic contrast-enhanced MRI in the brain.
Bourassa-Moreau, Benoît; Lebel, Réjean; Gilbert, Guillaume; Mathieu, David; Lepage, Martin.
Afiliação
  • Bourassa-Moreau B; Centre d'imagerie moléculaire de Sherbrooke, Département de médecine nucléaire et radiobiologie, Université de Sherbrooke, Sherbrooke, Québec, Canada.
  • Lebel R; Centre d'imagerie moléculaire de Sherbrooke, Département de médecine nucléaire et radiobiologie, Université de Sherbrooke, Sherbrooke, Québec, Canada.
  • Gilbert G; MR Clinical Science, Philips Healthcare Canada, Markham, Ontario, Canada.
  • Mathieu D; Service de neurochirurgie, Département de chirurgie, Université de Sherbrooke, Sherbrooke, Québec, Canada.
  • Lepage M; Centre de recherche du Centre hospitalier universitaire de Sherbrooke, Centre intégré de santé et de services sociaux de l'Estrie, Sherbrooke, Québec, Canada.
Magn Reson Med ; 86(6): 3052-3066, 2021 12.
Article em En | MEDLINE | ID: mdl-34268824
ABSTRACT

PURPOSE:

Accurately estimating the arterial input function for dynamic contrast-enhanced MRI is challenging. An arterial input function is typically determined from signal magnitude changes related to a contrast agent, often leading to underestimation of peak concentrations. Alternatively, signal phase recovers the accurate peak concentration for straight vessels but suffers from high noise. A recent method proposed to fit the signal in the complex plane by combining the advantages of the previous 2 methods. The purpose of this work is to refine this complex-based method to determine the venous output function (VOF), an arterial input function surrogate, from the superior sagittal sinus.

METHODS:

We propose a state-of-the-art complex-based method that includes direct compensation for blood inflow and signal phase correction accounting for the curvature of the superior sagittal sinus, generally assumed collinear with B0 . We compared the magnitude-, phase-, and complex-based VOF determination methods against various simulated biases as well as for 29 brain metastases patients.

RESULTS:

Angulation of the superior sagittal sinus relative to B0 varied widely within patients, and its effect on the signal phase caused an underestimation of peak concentrations of up to 65%. Correction significantly increased the VOF peak concentration for the phase- and complex-based VOFs in the cohort. The phase-based method recovered accurate peak concentrations but lacked precision in the tail of the VOF. Our complex-based VOF completely recovered the effect of inflow and resulted in a high-peak concentration with limited noise.

CONCLUSION:

The new complex-based method resulted in high-quality VOF robust against superior sagittal sinus curvature and variations in patient positioning.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Seio Sagital Superior Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Seio Sagital Superior Limite: Humans Idioma: En Ano de publicação: 2021 Tipo de documento: Article