Your browser doesn't support javascript.
loading
Ipsilateral and Simultaneous Comparison of Responses from Acceleromyography- and Electromyography-based Neuromuscular Monitors.
Nemes, Réka; Lengyel, Szabolcs; Nagy, György; Hampton, David R; Gray, Martyn; Renew, J Ross; Tassonyi, Edömér; Fülesdi, Béla; Brull, Sorin J.
Afiliação
  • Nemes R; From the Department of Anesthesiology and Intensive Care, University of Debrecen, Debrecen, Hungary.
  • Lengyel S; the Department of Tisza Research, Danube Research Institute, Centre for Ecological Research, Hungarian Academy of Sciences, Debrecen, Hungary.
  • Nagy G; From the Department of Anesthesiology and Intensive Care, University of Debrecen, Debrecen, Hungary.
  • Hampton DR; the Institute of Biotechnology, University of Cambridge, Cambridge, United Kingdom.
  • Gray M; Anaesthesia Technology Limited, Walshford, United Kingdom.
  • Renew JR; the Department of Anesthesiology and Perioperative Medicine, Mayo Clinic, Jacksonville, Florida.
  • Tassonyi E; From the Department of Anesthesiology and Intensive Care, University of Debrecen, Debrecen, Hungary.
  • Fülesdi B; From the Department of Anesthesiology and Intensive Care, University of Debrecen, Debrecen, Hungary.
  • Brull SJ; the Department of Anesthesiology and Perioperative Medicine, Mayo Clinic, Jacksonville, Florida.
Anesthesiology ; 135(4): 597-611, 2021 10 01.
Article em En | MEDLINE | ID: mdl-34329371
ABSTRACT

BACKGROUND:

The paucity of easy-to-use, reliable objective neuromuscular monitors is an obstacle to universal adoption of routine neuromuscular monitoring. Electromyography (EMG) has been proposed as the optimal neuromuscular monitoring technology since it addresses several acceleromyography limitations. This clinical study compared simultaneous neuromuscular responses recorded from induction of neuromuscular block until recovery using the acceleromyography-based TOF-Watch SX and EMG-based TetraGraph.

METHODS:

Fifty consenting patients participated. The acceleromyography and EMG devices analyzed simultaneous contractions (acceleromyography) and muscle action potentials (EMG) from the adductor pollicis muscle by synchronization via fiber optic cable link. Bland-Altman analysis described the agreement between devices during distinct phases of neuromuscular block. The primary endpoint was agreement of acceleromyography- and EMG-derived normalized train-of-four ratios greater than or equal to 80%. Secondary endpoints were agreement in the recovery train-of-four ratio range less than 80% and agreement of baseline train-of-four ratios between the devices.

RESULTS:

Acceleromyography showed normalized train-of-four ratio greater than or equal to 80% earlier than EMG. When acceleromyography showed train-of-four ratio greater than or equal to 80% (n = 2,929), the bias was 1.3 toward acceleromyography (limits of agreement, -14.0 to 16.6). When EMG showed train-of-four ratio greater than or equal to 80% (n = 2,284), the bias was -0.5 toward EMG (-14.7 to 13.6). In the acceleromyography range train-of-four ratio less than 80% (n = 2,802), the bias was 2.1 (-16.1 to 20.2), and in the EMG range train-of-four ratio less than 80% (n = 3,447), it was 2.6 (-14.4 to 19.6). Baseline train-of-four ratios were higher and more variable with acceleromyography than with EMG.

CONCLUSIONS:

Bias was lower than in previous studies. Limits of agreement were wider than expected because acceleromyography readings varied more than EMG both at baseline and during recovery. The EMG-based monitor had higher precision and greater repeatability than acceleromyography. This difference between monitors was even greater when EMG data were compared to raw (nonnormalized) acceleromyography measurements. The EMG monitor is a better indicator of adequate recovery from neuromuscular block and readiness for safe tracheal extubation than the acceleromyography monitor.
Assuntos

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Músculo Esquelético / Bloqueio Neuromuscular / Eletromiografia / Acelerometria / Monitorização Neurofisiológica Intraoperatória Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2021 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Músculo Esquelético / Bloqueio Neuromuscular / Eletromiografia / Acelerometria / Monitorização Neurofisiológica Intraoperatória Limite: Adult / Female / Humans / Male / Middle aged Idioma: En Ano de publicação: 2021 Tipo de documento: Article