Validation of the YuWell YE900 oscillometric blood pressure monitor for professional office use in adults and children according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018).
Zhang, Hui-Jie; Zhang, Juan; Wang, Su-Lan; Zhang, Jin; Teng, Li-Ni; Zhang, Shu-Juan; Zhou, De-Jun; Long, Ming-Zhi.
Afiliação
  • Zhang HJ; Clinical Trial Center.
  • Zhang J; Clinical Trial Center.
  • Wang SL; Department of Cardiovascular Medicine, The Second Affiliated Hospital of Nanjing Medical University, Nanjing, China.
  • Zhang J; Department of Cardiovascular Medicine, The Second Affiliated Hospital of Nanjing Medical University, Nanjing, China.
  • Teng LN; Department of Cardiovascular Medicine, The Second Affiliated Hospital of Nanjing Medical University, Nanjing, China.
  • Zhang SJ; Department of Cardiovascular Medicine, The Second Affiliated Hospital of Nanjing Medical University, Nanjing, China.
  • Zhou DJ; Department of Cardiovascular Medicine, The Second Affiliated Hospital of Nanjing Medical University, Nanjing, China.
  • Long MZ; Department of Cardiovascular Medicine, The Second Affiliated Hospital of Nanjing Medical University, Nanjing, China.
Blood Press Monit ; 26(5): 396-399, 2021 10 01.
Article em En | MEDLINE | ID: mdl-34480474
ABSTRACT

OBJECTIVE:

To evaluate the accuracy of the YuWell YE900 oscillometric upper-arm professional office blood pressure monitor in adults and children according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-22018).

METHODS:

Subjects were recruited to fulfill the age, sex, blood pressure and cuff distribution criteria of the AAMI/ESH/ISO Universal Standard in adults and children (aged 3-12 years) using the same arm sequential blood pressure measurement method. Three cuffs of the test device were used for arm circumference 18-22 cm (small), 22-32 cm (medium) and 32-42 cm (large).

RESULTS:

Ninety-two subjects were recruited, and 85 (50 adults and 35 children) were analyzed. For validation criterion 1, the mean ± SD of the differences between the test device and reference blood pressure readings was 1.7 ± 6.62/3.1 ± 5.76 mmHg (systolic/diastolic). For validation criterion 2, the SD of the averaged blood pressure differences between the test device and reference blood pressure per subject was 5.25/5.13 mmHg (systolic/diastolic).

CONCLUSION:

The YuWell YE900 professional electronic blood pressure monitor has passed the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-22018) in adults and children and can be recommended for clinical use.