Developing an ultra-performance liquid chromatography-tandem mass spectrometry for detecting aldosterone in human plasma.
Lin, Wenbin; Yao, Zhenrong; Li, Yuzhe; Liao, Zhihao; Xiao, Jiahao; Chen, Yonghong; Feng, Pinning; Gan, Wenjia.
Afiliação
  • Lin W; Department of Clinical Laboratory, The First Affiliated Hospital of Sun Yat-sen University, Guangzhou, China.
  • Yao Z; Department of Clinical Laboratory, The First Affiliated Hospital of Sun Yat-sen University, Guangzhou, China.
  • Li Y; Department of Clinical Laboratory, The Third Affiliated Hospital of Sun Yat-sen University, Guangzhou, China.
  • Liao Z; Department of Clinical Laboratory, The Third Affiliated Hospital of Sun Yat-sen University, Guangzhou, China.
  • Xiao J; Reproductive Medicine Center, The Third Affiliated Hospital of Sun Yat-sen University, Guangzhou, China.
  • Chen Y; Department of Clinical Laboratory, Sixth People's Hospital of Longgang District, Shenzhen, China.
  • Feng P; Department of Clinical Laboratory, The First Affiliated Hospital of Sun Yat-sen University, Guangzhou, China.
  • Gan W; Department of Clinical Laboratory, The First Affiliated Hospital of Sun Yat-sen University, Guangzhou, China.
J Clin Lab Anal ; 35(11): e24029, 2021 Nov.
Article em En | MEDLINE | ID: mdl-34590736
BACKGROUND: Accurately measuring plasma aldosterone concentration is difficult but meaningful for primary aldosteronism (PA) diagnosis. METHODS: In this study, we developed an ultra-performance liquid chromatography-tandem mass spectrometry (UPLC-MS/MS) method for plasma aldosterone detection, evaluated its performance according to guidelines issued by CLSI, including detection limit, linearity, precision, and compared it with chemiluminescence immunoassay. Then, a reference range of plasma aldosterone in young people was established by using this method. RESULTS: The lower limit of quantitation (LOQ) was 10 pg/ml. The mean recovery rates of analyte added to serum were 100.07-102.05% in different concentrations. The linearity range was 20-2000 pg/ml. Inter-assay CVs were 2.20-3.97% at aldosterone concentrations of 65.66-854.75 pg/ml. The regression equation of UPLC-MS/MS (x) and chemiluminescence immunoassay (y) was y = 1.002x + 65.854 (r = 0.9456, n = 237). The reference range of plasma aldosterone detected by UPLC-MS/MS was 11.30-363.82 pg/ml in young people in South China, and there was no statistically significant difference in plasma aldosterone concentration between two genders. CONCLUSION: In conclusion, UPLC-MS/MS can rapidly and accurately detect plasma aldosterone and is appropriate for clinical application.