In-vivo estimation of tissue electrical conductivities of a rabbit eye for precise simulation of electric field distributions during ocular iontophoresis.
Lee, Sangjun; Lee, Chany; Kim, Euijin; Ko, Song Ah; Kim, Se-Na; Choy, Young Bin; Im, Chang-Hwan.
Afiliação
  • Lee S; Department of Biomedical Engineering, Hanyang University, Seoul, Republic of Korea.
  • Lee C; Department of Electronic Engineering, Hanyang University, Seoul, Republic of Korea.
  • Kim E; Department of Structure & Function of Neural Network, Korea Brain Research Institute, Daegu, Republic of Korea.
  • Ko SA; Department of Biomedical Engineering, Hanyang University, Seoul, Republic of Korea.
  • Kim SN; Interdisciplinary Program in Bioengineering, College of Engineering, Seoul National University, Seoul, Republic of Korea.
  • Choy YB; Institute of Medical & Biological Engineering, Medical Research Center, Seoul National University, Seoul, Republic of Korea.
  • Im CH; Interdisciplinary Program in Bioengineering, College of Engineering, Seoul National University, Seoul, Republic of Korea.
Int J Numer Method Biomed Eng ; 38(1): e3540, 2022 01.
Article em En | MEDLINE | ID: mdl-34672120
ABSTRACT
Precise estimation of electrical conductivity of the eyes is important for the accurate analysis of electric field distributions in the eyes during ocular iontophoresis. In this study, we estimated the tissue electrical conductivities of a rabbit eye, which has been widely employed for neuro-ophthalmological experiments, through an in vivo experiment for the first time. Electrical potentials were measured at multiple locations on the skin, while weak currents were transmitted into the skin via two surface electrodes attached to the skin around the eye. A finite element model was constructed to calculate the electric potentials at the measurement locations. The conductivity values of different tissues were then estimated using an optimization procedure to minimize the difference between the measured and calculated electric potentials. The accuracy of the estimated tissue conductivity values of the rabbit eye was validated by comparing the measured and calculated electric potential values for different electrode montages. Further multi-physical analyses of iontophoretic drug delivery to the rabbit eye showed a significant influence of the conductivity profile on the resultant particle distribution. Overall, our results provide an important reference for the tissue electrical conductivity values of the rabbit eye, which could be further utilized for designing new medical devices for delivering electric fields to the eyes, such as transorbital and transscleral electrical stimulations.