Temperature profile and residual heat of monopolar laparoscopic and endoscopic dissection instruments.
Brinkmann, Franz; Hüttner, Ronny; Mehner, Philipp J; Henkel, Konrad; Paschew, Georgi; Herzog, Moritz; Martens, Nora; Richter, Andreas; Hinz, Sebastian; Groß, Justus; Schafmayer, Clemens; Hampe, Jochen; Hendricks, Alexander; Schwandner, Frank.
Afiliação
  • Brinkmann F; Department of Medicine I, University Hospital Dresden, Technische Universität Dresden (TU Dresden), Dresden, Germany. Franz.Brinkmann@uniklinikum-dresden.de.
  • Hüttner R; Chair of Microsystems, Faculty of Electrical and Computer Engineering, Technische Universität Dresden (TU Dresden), Dresden, Germany. Franz.Brinkmann@uniklinikum-dresden.de.
  • Mehner PJ; Else Kröner Fresenius Center for Digital Health, Technische Universität Dresden (TU Dresden), Dresden, Germany. Franz.Brinkmann@uniklinikum-dresden.de.
  • Henkel K; Department of Medicine I, University Hospital Carl Gustav Carus, Technische Universität Dresden (TU Dresden), Fetscherstr. 74, 01307, Dresden, Germany. Franz.Brinkmann@uniklinikum-dresden.de.
  • Paschew G; Chair of Microsystems, Faculty of Electrical and Computer Engineering, Technische Universität Dresden (TU Dresden), Dresden, Germany.
  • Herzog M; Chair of Microsystems, Faculty of Electrical and Computer Engineering, Technische Universität Dresden (TU Dresden), Dresden, Germany.
  • Martens N; Chair of Microsystems, Faculty of Electrical and Computer Engineering, Technische Universität Dresden (TU Dresden), Dresden, Germany.
  • Richter A; Chair of Microsystems, Faculty of Electrical and Computer Engineering, Technische Universität Dresden (TU Dresden), Dresden, Germany.
  • Hinz S; Department of Medicine I, University Hospital Dresden, Technische Universität Dresden (TU Dresden), Dresden, Germany.
  • Groß J; Else Kröner Fresenius Center for Digital Health, Technische Universität Dresden (TU Dresden), Dresden, Germany.
  • Schafmayer C; Department of Medicine I, University Hospital Dresden, Technische Universität Dresden (TU Dresden), Dresden, Germany.
  • Hampe J; Else Kröner Fresenius Center for Digital Health, Technische Universität Dresden (TU Dresden), Dresden, Germany.
  • Hendricks A; Chair of Microsystems, Faculty of Electrical and Computer Engineering, Technische Universität Dresden (TU Dresden), Dresden, Germany.
  • Schwandner F; Else Kröner Fresenius Center for Digital Health, Technische Universität Dresden (TU Dresden), Dresden, Germany.
Surg Endosc ; 36(6): 4507-4517, 2022 06.
Article em En | MEDLINE | ID: mdl-34708296
ABSTRACT

BACKGROUND:

Endoscopic and laparoscopic electrosurgical devices (ED) are of great importance in modern medicine but can cause adverse events such as tissue injuries and burns from residual heat. While laparoscopic tools are well investigated, detailed insights about the temperature profile of endoscopic knives are lacking. Our aim is to investigate the temperature and the residual heat of laparoscopic and endoscopic monopolar instruments to increase the safety in handling ED.

METHODS:

An infrared camera was used to measure the temperature of laparoscopic and endoscopic instruments during energy application and to determine the cooling time to below 50 °C at a porcine stomach. Different power levels and cutting intervals were studied to investigate their impact on the temperature profile.

RESULTS:

During activation, the laparoscopic hook exceeded 120 °C regularly for an up to 10 mm shaft length. With regards to endoknives, only the Dual Tip Knife showed a shaft temperature of above 50 °C. The residual heat of the laparoscopic hook remained above 50 °C for at least 15 s after activation. Endoknives cooled to below 50 °C in 4 s. A higher power level and longer cutting duration significantly increased the shaft temperature and prolonged the cooling time (p < 0.001).

CONCLUSION:

Residual heat and maximum temperature during energy application depend strongly on the chosen effect and cutting duration. To avoid potential injuries, the user should not touch any tissue with the laparoscopic hook for at least 15 s and with the endoknives for at least 4 s after energy application. As the shaft also heats up to over 120 °C, the user should be careful to avoid tissue contact during activation with the shaft. These results should be strongly considered for safety reasons when handling monopolar ED.