Dynamic MRI of swallowing: real-time volumetric imaging at 12 frames per second at 3 T.
Voskuilen, Luuk; Schoormans, Jasper; Gurney-Champion, Oliver J; Balm, Alfons J M; Strijkers, Gustav J; Smeele, Ludi E; Nederveen, Aart J.
Afiliação
  • Voskuilen L; Department of Head and Neck Oncology and Surgery, Netherlands Cancer Institute, Antoni van Leeuwenhoek, Plesmanlaan 121, 1066 CX, Amsterdam, The Netherlands. luukvoskuilen@gmail.com.
  • Schoormans J; Department of Radiology and Nuclear Medicine, Amsterdam University Medical Centers, University of Amsterdam, Cancer Center Amsterdam, Amsterdam, The Netherlands. luukvoskuilen@gmail.com.
  • Gurney-Champion OJ; Academic Centre for Dentistry Amsterdam and Academic Medical Center, University of Amsterdam and VU University Amsterdam, Amsterdam, The Netherlands. luukvoskuilen@gmail.com.
  • Balm AJM; Biomedical Engineering and Physics, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.
  • Strijkers GJ; Department of Radiology and Nuclear Medicine, Amsterdam University Medical Centers, University of Amsterdam, Cancer Center Amsterdam, Amsterdam, The Netherlands.
  • Smeele LE; Department of Head and Neck Oncology and Surgery, Netherlands Cancer Institute, Antoni van Leeuwenhoek, Plesmanlaan 121, 1066 CX, Amsterdam, The Netherlands.
  • Nederveen AJ; Department of Oral and Maxillofacial Surgery, Amsterdam University Medical Centers, University of Amsterdam, Amsterdam, The Netherlands.
MAGMA ; 35(3): 411-419, 2022 Jun.
Article em En | MEDLINE | ID: mdl-34779971
ABSTRACT

OBJECTIVE:

Dysphagia or difficulty in swallowing is a potentially hazardous clinical problem that needs regular monitoring. Real-time 2D MRI of swallowing is a promising radiation-free alternative to the current clinical standard videofluoroscopy. However, aspiration may be missed if it occurs outside this single imaged slice. We therefore aimed to image swallowing in 3D real time at 12 frames per second (fps). MATERIALS AND

METHODS:

At 3 T, three 3D real-time MRI acquisition approaches were compared to the 2D acquisition an aligned stack-of-stars (SOS), and a rotated SOS with a golden-angle increment and with a tiny golden-angle increment. The optimal 3D acquisition was determined by computer simulations and phantom scans. Subsequently, five healthy volunteers were scanned and swallowing parameters were measured.

RESULTS:

Although the rotated SOS approaches resulted in better image quality in simulations, in practice, the aligned SOS performed best due to the limited number of slices. The four swallowing phases could be distinguished in 3D real-time MRI, even though the spatial blurring was stronger than in 2D. The swallowing parameters were similar between 2 and 3D.

CONCLUSION:

At a spatial resolution of 2-by-2-by-6 mm with seven slices, swallowing can be imaged in 3D real time at a frame rate of 12 fps.