Zero-Fluoroscopy Cardiac Ablation: Technology Is Moving Forward in Complex Procedures-A Novel Workflow for Atrial Fibrillation.
Bertini, Matteo; Pompei, Graziella; Tolomeo, Paolo; Malagù, Michele; Fiorio, Alessio; Balla, Cristina; Vitali, Francesco; Rapezzi, Claudio.
Afiliação
  • Bertini M; Cardiologic Unit, Translational Medicine Department, Sant'Anna University Hospital of Ferrara, 44124 Ferrara, Italy.
  • Pompei G; Cardiologic Unit, Translational Medicine Department, Sant'Anna University Hospital of Ferrara, 44124 Ferrara, Italy.
  • Tolomeo P; Cardiologic Unit, Translational Medicine Department, Sant'Anna University Hospital of Ferrara, 44124 Ferrara, Italy.
  • Malagù M; Cardiologic Unit, Translational Medicine Department, Sant'Anna University Hospital of Ferrara, 44124 Ferrara, Italy.
  • Fiorio A; Cardiologic Unit, Translational Medicine Department, Sant'Anna University Hospital of Ferrara, 44124 Ferrara, Italy.
  • Balla C; Cardiologic Unit, Translational Medicine Department, Sant'Anna University Hospital of Ferrara, 44124 Ferrara, Italy.
  • Vitali F; Cardiologic Unit, Translational Medicine Department, Sant'Anna University Hospital of Ferrara, 44124 Ferrara, Italy.
  • Rapezzi C; Cardiologic Unit, Translational Medicine Department, Sant'Anna University Hospital of Ferrara, 44124 Ferrara, Italy.
Biology (Basel) ; 10(12)2021 Dec 15.
Article em En | MEDLINE | ID: mdl-34943247
ABSTRACT
BACKGROUND AND RATIONALE A fluoroscopy-based approach to an electrophysiological procedure is widely validated and has been recognized as the gold standard for a long time. The use of fluoroscopy exposes both the healthcare staff and the patient to a non-negligible dose of radiation. To minimize the risks associated with the use of fluoroscopy, it would be reasonable to perform ablation procedures with zero fluoroscopy. This approach is widely used in simple ablation procedures, but not in complex procedures. In atrial fibrillation (AF) ablation procedures, fluoroscopy remains the main technology used, in particular to guide the transseptal puncture. Main results and Implications. We present a workflow to perform a complete zero-fluoroscopy ablation for AF ablation procedures using a 3D electro-anatomical mapping system, intracardiac echocardiography and a novel steerable guiding sheath that can be visualized on the mapping system. We present two cases, one with paroxysmal AF and the other one with persistent AF during which we applied this novel workflow achieving a successful pulmonary vein isolation without complications and complete zero-fluoroscopy exposure.