Serological testing on the ADVIA Centaur system for human immunodeficiency virus, hepatitis B virus, and hepatitis C virus in specimens from deceased and living individuals demonstrates equivalent resultsƚ.
Wright, Theodore B; Patibandla, Sai; Walsh, Renee; Fonstad, Rachel; Gee, Matthew; Bitcon, Vera; Hopper, Julie; Braniff, Susie J; Best, Susan; Read, Scott.
Afiliação
  • Wright TB; Laboratory Diagnostics, Siemens Healthcare Diagnostics Inc., Tarrytown, New York, USA.
  • Patibandla S; Laboratory Diagnostics, Siemens Healthcare Diagnostics Inc., Tarrytown, New York, USA.
  • Walsh R; Laboratory Diagnostics, Siemens Healthcare Diagnostics Inc., Tarrytown, New York, USA.
  • Fonstad R; Laboratory Diagnostics, Siemens Healthcare Diagnostics Inc., Tarrytown, New York, USA.
  • Gee M; Laboratory Diagnostics, Siemens Healthcare Diagnostics Inc., Tarrytown, New York, USA.
  • Bitcon V; Laboratory Diagnostics, Siemens Healthcare Diagnostics Inc., Tarrytown, New York, USA.
  • Hopper J; Laboratory Diagnostics Inc., Siemens Healthcare Diagnostics Inc., Walpole, Massachusetts, USA.
  • Braniff SJ; National Serology Reference Laboratory, Fitzroy, Victoria, Australia.
  • Best S; National Serology Reference Laboratory, Fitzroy, Victoria, Australia.
  • Read S; Laboratory Diagnostics, Siemens Healthcare Pty Ltd., Bayswater, Victoria, Australia.
Transpl Infect Dis ; 24(2): e13802, 2022 Apr.
Article em En | MEDLINE | ID: mdl-35176197
ABSTRACT

INTRODUCTION:

To determine the suitability of human tissues and cells for transplantation, guidelines mandate infectious disease testing of serum or plasma obtained from deceased donors, which are often collected after cessation of the heartbeat. Tests used for this purpose are required to show equivalent performance when compared to pre-mortem specimens. This study evaluated whether serology assays for HIV Ag/Ab Combo, hepatitis B virus (HBc Total; HBsAgII), and HCV on the ADVIA Centaur system, were fit for testing post-mortem sera. Performance evaluation studies included precision, specificity, and sensitivity.

METHODS:

Blood specimens were collected within 24 h after death from 82 deceased and 83 healthy living individuals. Studies followed standard guidelines. The 20-day precision study was performed on five levels of post-mortem specimens (non-spiked and spiked). The specificity study compared 81-83 pre-mortem and 74-82 post-mortem specimens. The sensitivity study compared 50 pre-mortem and 50 post-mortem specimens spiked with positive sera for each analyte at two levels to achieve a low (near cutoff) positive result and a second higher positive result.

RESULTS:

Precision, specificity, and sensitivity study results met acceptance criteria for all assays and lots; post-mortem and pre-mortem results were equivalent.

CONCLUSION:

Based on this study, the ADVIA Centaur CHIV, HBcT, HBsAgII, and HCV assays are acceptable for use in routine testing of deceased donor sera collected after cessation of the heartbeat.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Infecções por HIV / Hepatite C / Hepatite B Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Infecções por HIV / Hepatite C / Hepatite B Tipo de estudo: Diagnostic_studies Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article