A Comparative Study of Birdcage RF Coil Configurations for Ultra-High Field Magnetic Resonance Imaging.
Sensors (Basel)
; 22(5)2022 Feb 23.
Article
em En
| MEDLINE
| ID: mdl-35270889
Improvements in transmission and reception sensitivities of radiofrequency (RF) coils used in ultra-high field (UHF) magnetic resonance imaging (MRI) are needed to reduce specific absorption rates (SAR) and RF power deposition, albeit without applying high-power RF. Here, we propose a method to simultaneously improve transmission efficiency and reception sensitivity of a band-pass birdcage RF coil (BP-BC RF coil) by combining a multi-channel wireless RF element (MCWE) with a high permittivity material (HPM) in a 7.0 T MRI. Electromagnetic field (EM-field) simulations, performed using two types of phantoms, viz., a cylindrical phantom filled with oil and a human head model, were used to compare the effects of MCWE and HPM on BP-BC RF coils. EM-fields were calculated using the finite difference time-domain (FDTD) method and analyzed using Matlab software. Next, to improve RF transmission efficiency, we compared two HPM structures, namely, a hollow cylinder shape HPM (hcHPM) and segmented cylinder shape HPM (scHPM). The scHPM and MCWE model comprised 16 elements (16-rad BP-BC RF coil) and this coil configuration demonstrated superior RF transmission efficiency and reception sensitivity along with an acceptable SAR. We expect wider clinical application of this combination in 7.0 T MRIs, which were recently approved by the United States Food and Drug Administration.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Ondas de Rádio
/
Imageamento por Ressonância Magnética
Limite:
Humans
País como assunto:
America do norte
Idioma:
En
Ano de publicação:
2022
Tipo de documento:
Article