Performance evaluation of UF-4000 body fluid mode for automated body fluid cell counting.
Siatkowski, Marie; Dahyot, Sandrine; Pestel-Caron, Martine; Boyer, Sophie.
Afiliação
  • Siatkowski M; CHU Rouen, Department of Microbiology, F-76000 Rouen, France.
  • Dahyot S; Normandie Univ, UNIROUEN, UNICAEN, Inserm U 1311, CHU Rouen, Department of Microbiology, F-76000 Rouen, France.
  • Pestel-Caron M; Normandie Univ, UNIROUEN, UNICAEN, Inserm U 1311, CHU Rouen, Department of Microbiology, F-76000 Rouen, France.
  • Boyer S; Normandie Univ, UNIROUEN, UNICAEN, Inserm U 1311, CHU Rouen, Department of Microbiology, F-76000 Rouen, France. Electronic address: sophie.boyer@chu-rouen.fr.
Clin Chim Acta ; 531: 152-156, 2022 Jun 01.
Article em En | MEDLINE | ID: mdl-35398022
ABSTRACT

BACKGROUND:

Cytological analysis of body fluids (BF) provides important information for diagnosis in various medical conditions. We evaluated the analytical performance of the UF-4000 BF mode for ascitic, cerebrospinal, pleural, synovial and continuous ambulatory peritoneal dialysis fluids compared to light microscopy counting (LM). MATERIALS AND

METHODS:

223 consecutive BF were analyzed by UF-4000 and results were compared using Pearson's correlation, Bland-Altman analysis, and contingence tests at relevant cut-off values. This study also included the evaluation of precision, linearity, and carryover.

RESULTS:

For white and red blood cells (WBC, RBC) counts in all BF, correlation was excellent with Pearson's coefficients R2 > 0,98. Bland-Altman analysis didn't reveal significant differences with limited bias for WBC ranging from -10 to -1 WBC/µL and bias ranging from -43 to -6/µL for RBC. At specific cut-off values for WBC, Se and Spe were 100% except for ascites (Spe = 98%) due to two false positive. Precision evaluated at three concentration levels was good for each parameter (WBC < 10%). Linearity was excellent for WBC (R2 > 0,99) and carryover negligible (<0,004%).

CONCLUSION:

UF-4000 BF mode is a good alternative to manual LM for BF cell counting. This automated method gives rapid and accurate results which is important for therapeutic decisions.