Radiation Dose of Contrast-Enhanced Mammography: A Two-Center Prospective Comparison.
Gennaro, Gisella; Cozzi, Andrea; Schiaffino, Simone; Sardanelli, Francesco; Caumo, Francesca.
Afiliação
  • Gennaro G; Unit of Breast Radiology, Veneto Institute of Oncology (IOV) IRCCS, Via Gattamelata 64, 35128 Padua, Italy.
  • Cozzi A; Department of Biomedical Sciences for Health, Università degli Studi di Milano, Via Mangiagalli 31, 20133 Milano, Italy.
  • Schiaffino S; Unit of Radiology, IRCCS Policlinico San Donato, Via Morandi 30, 20097 San Donato Milanese, Italy.
  • Sardanelli F; Department of Biomedical Sciences for Health, Università degli Studi di Milano, Via Mangiagalli 31, 20133 Milano, Italy.
  • Caumo F; Unit of Radiology, IRCCS Policlinico San Donato, Via Morandi 30, 20097 San Donato Milanese, Italy.
Cancers (Basel) ; 14(7)2022 Mar 31.
Article em En | MEDLINE | ID: mdl-35406546
ABSTRACT
The radiation dose associated with contrast-enhanced mammography (CEM) has been investigated only by single-center studies. In this retrospective study, we aimed to compare the radiation dose between two centers performing CEM within two prospective studies, using the same type of equipment. The CEM mean glandular dose (MGD) was computed for low energy (LE) and high energy (HE) images and their sum was calculated for each view. MGD and related parameters (entrance dose, breast thickness, compression, and density) were compared between the two centers using the Mann−Whitney test. Finally, per-patient MGD was calculated by pooling the two datasets and determining the contribution of LE and HE images. A total of 348 CEM examinations were analyzed (228 from Center 1 and 120 from Center 2). The median total MGD per view was 2.33 mGy (interquartile range 2.19−2.51 mGy) at Center 1 and 2.46 mGy (interquartile range 2.32−2.70 mGy) at Center 2, with a 0.15 mGy median difference (p < 0.001) equal to 6.2%. LE-images contributed between 64% and 77% to the total patient dose in CEM, with the remaining 23−36% being associated with HE images. The mean radiation dose for a two-view bilateral CEM exam was 4.90 mGy, about 30% higher than for digital mammography.