Your browser doesn't support javascript.
loading
Investigation of the Accuracy of Four Intraoral Scanners in Mandibular Full-Arch Digital Implant Impression: A Comparative In Vitro Study.
Di Fiore, Adolfo; Graiff, Lorenzo; Savio, Gianpaolo; Granata, Stefano; Basilicata, Michele; Bollero, Patrizio; Meneghello, Roberto.
Afiliação
  • Di Fiore A; Department of Neurosciences, Section of Prosthetic and Digital Dentistry, University of Padova, 35100 Padova, Italy.
  • Graiff L; Department of Neurosciences, Section of Prosthetic and Digital Dentistry, University of Padova, 35100 Padova, Italy.
  • Savio G; Departments of Civil, Environmental and Architectural Engineering, University of Padova, 35100 Padova, Italy.
  • Granata S; Department of Neurosciences, Section of Prosthetic and Digital Dentistry, University of Padova, 35100 Padova, Italy.
  • Basilicata M; Department of Systems Medicine, University of Rome Tor Vergata, 00133 Roma, Italy.
  • Bollero P; Department of Systems Medicine, University of Rome Tor Vergata, 00133 Roma, Italy.
  • Meneghello R; Department of Management and Engineering, University of Padova, 35100 Padova, Italy.
Article em En | MEDLINE | ID: mdl-35457583
ABSTRACT

Background:

We compare the accuracy of new intraoral scanners (IOSs) in full-arch digital implant impressions.

Methods:

A master model with six scan bodies was milled in poly(methyl methacrylate), measured by using a coordinate measuring machine, and scanned 15 times with four IOSs PrimeScan, Medit i500, Vatech EZ scan, and iTero. The software was developed to identify the position points on each scan body. The 3D position and distance analysis were performed.

Results:

The average and ± standard deviation of the 3D position analysis was 29 µm ± 6 µm for PrimeScan, 39 µm ± 6 µm for iTero, 48 µm ± 18 µm for Mediti500, and 118 µm ± 24 µm for Vatech EZ scan (p < 0.05).

Conclusions:

All IOSs are able to make a digital complete implant impression in vitro according to the average misfit value reported in literature (150 µm); however, the 3D distance analysis showed that only the Primescan and iTero presented negligible systematic error sources.