Systolic blood pressure measurements are unreliable for the management of acute spontaneous intracerebral hemorrhage.
J Crit Care
; 70: 154049, 2022 08.
Article
em En
| MEDLINE
| ID: mdl-35490501
PURPOSE: Whether systolic blood pressure (SBP) is reliable in acute spontaneous intracerebral (sICH) by assessing agreement between simultaneous BP measurements obtained from cuff non-invasive blood pressure (NIBP) and radial arterial invasive blood pressure (AIBP) devices. MATERIAL AND METHODS: Among 766 prospectively screened sICH subjects, 303 (39.5%) had NIBP and AIBP measurements. During the first 24 h, 2157 simultaneous paired measurement readings were abstracted. Paired NIBP/AIBP measurements were included in a Bland-Altman technique with 95% agreement limits and coefficients from regression analysis derived from a bootstrap procedure. RESULTS: Variance for SBP was 66.1 mmHg, which was larger than the 44.3 mg Hg for diastolic blood pressure (DBP) or the 46.1 mmHg for mean arterial pressure (MAP). Pairwise comparison of mean biases showed a significant difference between SBP when compared to DBP (p < 0.0001) or MAP (p < 0.0001). The mean bias between DBP and MAP was not different (p = 0.68). Regression-based Bland Altman analysis found significant bias (slope -0.16, 95% CI -0.23, -0.09, p < 0.05) over the range of mean SBP. Bias over the range of mean DBP or MAP was not significant. CONCLUSIONS: We concluded that SBP is an unreliable blood pressure measurement in patients with sICH.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Determinação da Pressão Arterial
/
Pressão Arterial
Tipo de estudo:
Diagnostic_studies
Limite:
Humans
Idioma:
En
Ano de publicação:
2022
Tipo de documento:
Article