Increasing angular sampling through deep learning for stationary cardiac SPECT image reconstruction.
Xie, Huidong; Thorn, Stephanie; Chen, Xiongchao; Zhou, Bo; Liu, Hui; Liu, Zhao; Lee, Supum; Wang, Ge; Liu, Yi-Hwa; Sinusas, Albert J; Liu, Chi.
Afiliação
  • Xie H; Department of Biomedical Engineering, Yale University, 801 Howard Avenue, New Haven, CT, 06520, USA.
  • Thorn S; Department of Internal Medicine (Cardiology), Yale University, New Haven, CT, USA.
  • Chen X; Department of Biomedical Engineering, Yale University, 801 Howard Avenue, New Haven, CT, 06520, USA.
  • Zhou B; Department of Biomedical Engineering, Yale University, 801 Howard Avenue, New Haven, CT, 06520, USA.
  • Liu H; Department of Radiology and Biomedical Imaging, Yale University, 801 Howard Avenue, New Haven, CT, 06520, USA.
  • Liu Z; Department of Engineering Physics, Tsinghua University, Beijing, China.
  • Lee S; Department of Radiology and Biomedical Imaging, Yale University, 801 Howard Avenue, New Haven, CT, 06520, USA.
  • Wang G; Department of Internal Medicine (Cardiology), Yale University, New Haven, CT, USA.
  • Liu YH; Department of Biomedical Engineering, Rensselaer Polytechnic Institute, Troy, NY, USA.
  • Sinusas AJ; Department of Internal Medicine (Cardiology), Yale University, New Haven, CT, USA.
  • Liu C; Department of Biomedical Imaging and Radiological Sciences, School of Biomedical Science and Engineering, National Yang Ming Chiao Tung University, Taipei, Taiwan.
J Nucl Cardiol ; 30(1): 86-100, 2023 02.
Article em En | MEDLINE | ID: mdl-35508796
ABSTRACT

BACKGROUND:

The GE Discovery NM (DNM) 530c/570c are dedicated cardiac SPECT scanners with 19 detector modules designed for stationary imaging. This study aims to incorporate additional projection angular sampling to improve reconstruction quality. A deep learning method is also proposed to generate synthetic dense-view image volumes from few-view counterparts.

METHODS:

By moving the detector array, a total of four projection angle sets were acquired and combined for image reconstructions. A deep neural network is proposed to generate synthetic four-angle images with 76 ([Formula see text]) projections from corresponding one-angle images with 19 projections. Simulated data, pig, physical phantom, and human studies were used for network training and evaluation. Reconstruction results were quantitatively evaluated using representative image metrics. The myocardial perfusion defect size of different subjects was quantified using an FDA-cleared clinical software.

RESULTS:

Multi-angle reconstructions and network results have higher image resolution, improved uniformity on normal myocardium, more accurate defect quantification, and superior quantitative values on all the testing data. As validated against cardiac catheterization and diagnostic results, deep learning results showed improved image quality with better defect contrast on human studies.

CONCLUSION:

Increasing angular sampling can substantially improve image quality on DNM, and deep learning can be implemented to improve reconstruction quality in case of stationary imaging.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aprendizado Profundo Limite: Animals / Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Aprendizado Profundo Limite: Animals / Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article