Your browser doesn't support javascript.
loading
Free-breathing motion-informed locally low-rank quantitative 3D myocardial perfusion imaging.
Hoh, Tobias; Vishnevskiy, Valery; Polacin, Malgorzata; Manka, Robert; Fuetterer, Maximilian; Kozerke, Sebastian.
Afiliação
  • Hoh T; Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.
  • Vishnevskiy V; Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.
  • Polacin M; Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.
  • Manka R; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
  • Fuetterer M; Institute for Biomedical Engineering, University and ETH Zurich, Zurich, Switzerland.
  • Kozerke S; Institute of Diagnostic and Interventional Radiology, University Hospital Zurich, University of Zurich, Zurich, Switzerland.
Magn Reson Med ; 88(4): 1575-1591, 2022 10.
Article em En | MEDLINE | ID: mdl-35713206
ABSTRACT

PURPOSE:

To propose respiratory motion-informed locally low-rank reconstruction (MI-LLR) for robust free-breathing single-bolus quantitative 3D myocardial perfusion CMR imaging. Simulation and in-vivo results are compared to locally low-rank (LLR) and compressed sensing reconstructions (CS) for reference.

METHODS:

Data were acquired using a 3D Cartesian pseudo-spiral in-out k-t undersampling scheme (R = 10) and reconstructed using MI-LLR, which encompasses two stages. In the first stage, approximate displacement fields are derived from an initial LLR reconstruction to feed a motion-compensated reference system to a second reconstruction stage, which reduces the rank of the inverse problem. For comparison, data were also reconstructed with LLR and frame-by-frame CS using wavelets as sparsifying transform ( ℓ1$$ {\ell}_1 $$ -wavelet). Reconstruction accuracy relative to ground truth was assessed using synthetic data for realistic ranges of breathing motion, heart rates, and SNRs. In-vivo experiments were conducted in healthy subjects at rest and during adenosine stress. Myocardial blood flow (MBF) maps were derived using a Fermi model.

RESULTS:

Improved uniformity of MBF maps with reduced local variations was achieved with MI-LLR. For rest and stress, intra-volunteer variation of absolute and relative MBF was lower in MI-LLR (±0.17 mL/g/min [26%] and ±1.07 mL/g/min [33%]) versus LLR (±0.19 mL/g/min [28%] and ±1.22 mL/g/min [36%]) and versus ℓ1$$ {\ell}_1 $$ -wavelet (±1.17 mL/g/min [113%] and ±6.87 mL/g/min [115%]). At rest, intra-subject MBF variation was reduced significantly with MI-LLR.

CONCLUSION:

The combination of pseudo-spiral Cartesian undersampling and dual-stage MI-LLR reconstruction improves free-breathing quantitative 3D myocardial perfusion CMR imaging under rest and stress condition.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imagem de Perfusão do Miocárdio Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Imagem de Perfusão do Miocárdio Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article