Patient tilt improves efficacy of CO2 field-flooding in minimally invasive cardiac surgery.
Vandenberghe, Stijn; Singjeli, Geni; Demertzis, Stefanos.
Afiliação
  • Vandenberghe S; Faculty of Biomedical Sciences, Università Della Svizzera Italiana, Via Giuseppe Buffi 13, 6900, Lugano, Switzerland. Stijn.vandenberghe@usi.ch.
  • Singjeli G; Department of Cardiac Surgery, Istituto Cardiocentro Ticino, EOC, Lugano, Switzerland. Stijn.vandenberghe@usi.ch.
  • Demertzis S; Department of Cardiac Surgery, Istituto Cardiocentro Ticino, EOC, Lugano, Switzerland.
J Cardiothorac Surg ; 17(1): 164, 2022 Jun 22.
Article em En | MEDLINE | ID: mdl-35733212
ABSTRACT

OBJECTIVE:

Space limitations during minimally invasive cardiac surgery impede consistent use of CO2 field-flooding. We compared different gas delivery methods, flow rates and the effect of patient inclination.

METHODS:

A gastight model of MICS surgery with internal organs and right thoracotomy wound was created from a mannequin and equipped with a CO2 concentration sensor in the left ventricle. Maximum achievable CO2 concentration was compared for gas delivery via three commercial CO2 diffusors (CarbonMini, Temed, Andocor) and also via a trocar with side port. Gas flow rates of 1, 3, 5 and 8 L per minute were tested. The model was placed either in supine position or with 20° oblique tilt. A simplified transparent model was also created and placed in an optical test bench to evaluate the gas cloud motions via real-time visualization.

RESULTS:

The trocar consistently achieved higher CO2 concentrations inside the left ventricle. At 1 l/min, approximately 2.5 min were needed to fill the supine model to its maximum CO2 concentration, which was limited to a range of 48-82% in the left ventricle. At higher flow rates, filling time and concentration were significantly improved. In a tilted model, all devices and all flow rates generated on average 99% CO2 in the ventricle. Imaging revealed constant gas exchange via the main incision, with CO2 outflow via bottom and air inflow via the top of the incision.

CONCLUSIONS:

CO2 field flooding in minimally invasive cardiac surgery is highly effective if the patient is tilted. Else a flow rate of 5 l/min is recommended to achieve the same protection.