Low-Dose High-Resolution Photon-Counting CT of the Lung: Radiation Dose and Image Quality in the Clinical Routine.
Woeltjen, Matthias Michael; Niehoff, Julius Henning; Michael, Arwed Elias; Horstmeier, Sebastian; Moenninghoff, Christoph; Borggrefe, Jan; Kroeger, Jan Robert.
Afiliação
  • Woeltjen MM; Department of Radiology, Neuroradiology and Nuclear Medicine, Johannes Wesling University Hospital, Ruhr University Bochum, 44801 Bochum, Germany.
  • Niehoff JH; Department of Radiology, Neuroradiology and Nuclear Medicine, Johannes Wesling University Hospital, Ruhr University Bochum, 44801 Bochum, Germany.
  • Michael AE; Department of Radiology, Neuroradiology and Nuclear Medicine, Johannes Wesling University Hospital, Ruhr University Bochum, 44801 Bochum, Germany.
  • Horstmeier S; Department of Radiology, Neuroradiology and Nuclear Medicine, Johannes Wesling University Hospital, Ruhr University Bochum, 44801 Bochum, Germany.
  • Moenninghoff C; Department of Radiology, Neuroradiology and Nuclear Medicine, Johannes Wesling University Hospital, Ruhr University Bochum, 44801 Bochum, Germany.
  • Borggrefe J; Department of Radiology, Neuroradiology and Nuclear Medicine, Johannes Wesling University Hospital, Ruhr University Bochum, 44801 Bochum, Germany.
  • Kroeger JR; Department of Radiology, Neuroradiology and Nuclear Medicine, Johannes Wesling University Hospital, Ruhr University Bochum, 44801 Bochum, Germany.
Diagnostics (Basel) ; 12(6)2022 Jun 11.
Article em En | MEDLINE | ID: mdl-35741251
ABSTRACT
This study aims to investigate the qualitative and quantitative image quality of low-dose high-resolution (LD-HR) lung CT scans acquired with the first clinical approved photon counting CT (PCCT) scanner. Furthermore, the radiation dose used by the PCCT is compared to a conventional CT scanner with an energy-integrating detector system (EID-CT). Twenty-nine patients who underwent a LD-HR chest CT scan with dual-source PCCT and had previously undergone a LD-HR chest CT with a standard EID-CT scanner were retrospectively included in this study. Images of the whole lung as well as enlarged image sections displaying a specific finding (lesion) were evaluated in terms of overall image quality, image sharpness and image noise by three senior radiologists using a 5-point Likert scale. The PCCT images were reconstructed with and without a quantum iterative reconstruction algorithm (PCCT QIR+/−). Noise and signal-to-noise (SNR) were measured and the effective radiation dose was calculated. Overall, image quality and image sharpness were rated best in PCCT (QIR+) images. A significant difference was seen particularly in image sections of PCCT (QIR+) images compared to EID-CT images (p < 0.005). Image noise of PCCT (QIR+) images was significantly lower compared to EID-CT images in image sections (p = 0.005). In contrast, noise was lowest on EID-CT images (p < 0.001). The PCCT used significantly less radiation dose compared to the EID-CT (p < 0.001). In conclusion, LD-HR PCCT scans of the lung provide better image quality while using significantly less radiation dose compared to EID-CT scans.