In vitro accuracy of digital and conventional impressions in the partially edentulous maxilla.
Waldecker, Moritz; Rues, Stefan; Awounvo Awounvo, Junior Sinclair; Rammelsberg, Peter; Bömicke, Wolfgang.
Afiliação
  • Waldecker M; Department of Prosthetic Dentistry, Heidelberg University Hospital, University of Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany. moritz.waldecker@med.uni-heidelberg.de.
  • Rues S; Department of Prosthetic Dentistry, Heidelberg University Hospital, University of Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany.
  • Awounvo Awounvo JS; Institute of Medical Biometry, Heidelberg University Hospital, University of Heidelberg, Heidelberg, Germany.
  • Rammelsberg P; Department of Prosthetic Dentistry, Heidelberg University Hospital, University of Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany.
  • Bömicke W; Department of Prosthetic Dentistry, Heidelberg University Hospital, University of Heidelberg, Im Neuenheimer Feld 400, 69120, Heidelberg, Germany.
Clin Oral Investig ; 26(11): 6491-6502, 2022 Nov.
Article em En | MEDLINE | ID: mdl-35778534
OBJECTIVES: This in vitro study compared the dimensional accuracy of conventional impressions (CI) with that of digital impressions (DI) in a partially edentulous maxilla. DIs were made by two intraoral scanners, Omnicam (OC) and Primescan (PS). MATERIALS AND METHODS: CI and both intraoral scanners were used to take 30 impressions of two identical reference models. CIs were poured with type 4 gypsum and the saw-cut models were digitized. The reference models simulated a maxilla with six prepared teeth that accommodated a cross-arch fixed partial denture. Center points of five precision balls and center points at the margin level of each prepared tooth were used to detect changes in dimensions and tooth axis between the reference model and the scans. RESULTS: For DI, the largest deviations (176 µm for OC and 122 µm for PS) occurred over the cross-arch. For CI, the largest deviation (118 µm) occurred over the anterior segment. For shorter distances up to a quadrant, DI was superior to CI. For longer scan distances, DI was comparable (2 sextant and anterior segment) or inferior (cross-arch) to CI. Vertical and tooth axis deviations were significantly smaller for CI than for DI (p < 0.001). CONCLUSIONS: The impression method affected the impression accuracy of a partially edentulous maxilla with prepared teeth. DI is recommended for scans up to a quadrant. Larger scan volumes are not yet suitable for fabricating a fixed partial denture because of the high scatter of accuracy values. CLINICAL RELEVANCE: In contrast to conventional impressions, digital impressions lead to comparable or better results concerning scans up to a quadrant. Consequently, for larger scan volumes, several smaller scans should be performed or, if restoration-related not possible, it is recommended to take conventional impressions.