Impurities in commercial titanium dental implants - A mass and optical emission spectrometry elemental analysis.
Stricker, Andres; Bergfeldt, Thomas; Fretwurst, Tobias; Addison, Owen; Schmelzeisen, Rainer; Rothweiler, René; Nelson, Katja; Gross, Christian.
Afiliação
  • Stricker A; Department of Oral and Maxillofacial Surgery, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Hugstetter Str. 55, 79106 Freiburg im Breisgau, Germany. Electronic address: andres.stricker@uniklinik-freiburg.de.
  • Bergfeldt T; Institute of Applied Materials - Applied Material Physics (IAM-AWP), Karlsruhe Institute of Technology (KIT), Hermann-von-Helmholtz-Platz 1, 76344 Eggenstein-Leopoldshafen, Germany. Electronic address: thomas.bergfeldt@kit.edu.
  • Fretwurst T; Department of Oral and Maxillofacial Surgery, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Hugstetter Str. 55, 79106 Freiburg im Breisgau, Germany. Electronic address: tobias.fretwurst@uniklinik-freiburg.de.
  • Addison O; Centre for Oral, Clinical & Translational Sciences, Faculty of Dentistry, Oral and Craniofacial Sciences, King's College London, Guy's Hospital, Great Maze Pond, SE1 9RT London, UK. Electronic address: owen.addison@kcl.ac.uk.
  • Schmelzeisen R; Department of Oral and Maxillofacial Surgery, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Hugstetter Str. 55, 79106 Freiburg im Breisgau, Germany. Electronic address: rainer.schmelzeisen@uniklinik-freiburg.de.
  • Rothweiler R; Department of Oral and Maxillofacial Surgery, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Hugstetter Str. 55, 79106 Freiburg im Breisgau, Germany. Electronic address: rene.rothweiler@uniklinik-freiburg.de.
  • Nelson K; Department of Oral and Maxillofacial Surgery, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Hugstetter Str. 55, 79106 Freiburg im Breisgau, Germany. Electronic address: katja.nelson@uniklinik-freiburg.de.
  • Gross C; Department of Oral and Maxillofacial Surgery, Medical Center - University of Freiburg, Faculty of Medicine, University of Freiburg, Hugstetter Str. 55, 79106 Freiburg im Breisgau, Germany. Electronic address: christian.gross@uniklinik-freiburg.de.
Dent Mater ; 38(8): 1395-1403, 2022 08.
Article em En | MEDLINE | ID: mdl-35781168
ABSTRACT

OBJECTIVE:

Titanium (Ti) is considered bioinert and is still regarded as the "gold standard" material for dental implants. However, even 'commercial pure' Ti will contain minor fractions of elemental impurities. Evidence demonstrating the release of Ti ions and particles from 'passive' implant surfaces is increasing and has been attributed to biocorrosion processes which may provoke immunological reactions. However, Ti observed in peri-implant tissues has been shown to be co-located with elements considered impurities in biomedical alloys. Accordingly, this study aimed to quantify the composition of impurities in commercial Ti dental implants.

METHODS:

Fifteen commercial titanium dental implant systems were analyzed using inductively coupled plasma-mass spectrometry (ICP-MS) and optical emission spectrometry (ICP-OES).

RESULTS:

The elemental composition of implants manufactured from commercially pure grades of Ti, Ti-6Al-4V, and the TiZr alloy (Roxolid) conformed to the respective ISO/ASTM standards or manufacturers´ data (TiZr/Roxolid). However, all implants investigated included exogenous metal contaminants including Ni, Cr, Sb, and Nb to a variable extent. Other contaminants detected in a fraction of implants included As and the radionuclides U-238 and Th-232.

SIGNIFICANCE:

Although all Ti implant studies conformed with their standard compositions, potentially allergenic, noxious metals and even radionuclides were detected. Since there are differences in the degree of contamination between the implant systems, a certain impurity fraction seems technically avoidable. The clinical relevance of these findings must be further investigated, and an adaptation of industry standards should be discussed.