Six-lead device superior to single-lead smartwatch ECG in atrial fibrillation detection.
Scholten, Josca; Jansen, Ward P J; Horsthuis, Thomas; Mahes, Anuska D; Winter, Michiel M; Zwinderman, Aeilko H; Keijer, Jan T; Minneboo, Madelon; de Groot, Joris R; Bokma, Jouke P.
Afiliação
  • Scholten J; Department of Cardiology, Tergooi Hospital, Blaricum, North Holland, the Netherlands; Location Academic Medical Center, Department of Cardiology, Amsterdam University Medical Center, Meibergdreef 9, Amsterdam, North Holland 1105 AZ, the Netherlands.
  • Jansen WPJ; Department of Cardiology, Tergooi Hospital, Blaricum, North Holland, the Netherlands.
  • Horsthuis T; Department of Cardiology, Tergooi Hospital, Blaricum, North Holland, the Netherlands.
  • Mahes AD; Department of Cardiology, Tergooi Hospital, Blaricum, North Holland, the Netherlands.
  • Winter MM; Location Academic Medical Center, Department of Cardiology, Amsterdam University Medical Center, Meibergdreef 9, Amsterdam, North Holland 1105 AZ, the Netherlands.
  • Zwinderman AH; Location Academic Medical Center, Department of Clinical Epidemiology, Biostatistics and Bioinformatics, Amsterdam University Medical Center, Amsterdam, North Holland, the Netherlands.
  • Keijer JT; Department of Cardiology, Tergooi Hospital, Blaricum, North Holland, the Netherlands.
  • Minneboo M; Department of Cardiology, Tergooi Hospital, Blaricum, North Holland, the Netherlands; Location Academic Medical Center, Department of Cardiology, Amsterdam University Medical Center, Meibergdreef 9, Amsterdam, North Holland 1105 AZ, the Netherlands.
  • de Groot JR; Location Academic Medical Center, Department of Cardiology, Amsterdam University Medical Center, Meibergdreef 9, Amsterdam, North Holland 1105 AZ, the Netherlands.
  • Bokma JP; Department of Cardiology, Tergooi Hospital, Blaricum, North Holland, the Netherlands; Location Academic Medical Center, Department of Cardiology, Amsterdam University Medical Center, Meibergdreef 9, Amsterdam, North Holland 1105 AZ, the Netherlands. Electronic address: j.p.bokma@amsterdamumc.nl.
Am Heart J ; 253: 53-58, 2022 11.
Article em En | MEDLINE | ID: mdl-35850242
ABSTRACT
This was a head-to-head comparative study on different electrocardiogram (ECG)-based smartwatches and devices for atrial fibrillation detection. We prospectively included 220 patients scheduled for electrical cardioversion and recorded ECGs with 3 different devices (Withings Move ECG, Apple Watch 5, Kardia Mobile 6-leads) as well as the standard 12-lead ECG (gold standard), both before and after cardioversion. All atrial fibrillation detection algorithms had high accuracy (sensitivity and specificity 91-99%) but were hampered by uninterpretable recordings (20-24%). In cardiologists' interpretation, the 6-lead device was superior (sensitivity 99%, specificity 97%) to both single-lead smartwatches (P < .05) for atrial fibrillation detection.