Performance evaluation of presepsin using a Sysmex HISCL-5000 analyzer and determination of reference interval.
Kang, Taewon; Yoo, Jeaeun; Choi, Hyunyu; Lee, Seungok; Jekarl, Dong Wook; Kim, Yonggoo.
Afiliação
  • Kang T; Departement of Laboratory Medicine, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, South Korea.
  • Yoo J; Research and Development Institute for In Vitro Diagnostic Medical Devices of Catholic University of Korea, College of Medicine, The Catholic University of Korea, Seoul, South Korea.
  • Choi H; Departement of Laboratory Medicine, Incheon St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, South Korea.
  • Lee S; Departement of Laboratory Medicine, Incheon St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, South Korea.
  • Jekarl DW; Departement of Laboratory Medicine, Incheon St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, South Korea.
  • Kim Y; Departement of Laboratory Medicine, Seoul St. Mary's Hospital, College of Medicine, The Catholic University of Korea, Seoul, South Korea.
J Clin Lab Anal ; 36(9): e24618, 2022 Sep.
Article em En | MEDLINE | ID: mdl-35870180
ABSTRACT

BACKGROUND:

Analytical evaluation of newly developed presepsin by a Sysmex HISCL-5000 (Sysmex, Japan) automated immune analyzer was performed.

METHODS:

For evaluation, sepsis patient samples were collected before treatment in an emergency department. Precision, linearity, limit of blank/limit of detection, method comparisons, and reference intervals were evaluated. Method comparisons were performed using a PATHFAST immune analyzer (LSI Medience Corporation, Japan).

RESULTS:

Precision using a 20x2x2 protocol for low (306 pg/mL) and high (1031 pg/mL) levels resulted in within-laboratory standard deviation (95% confidence interval [CI]) and coefficient of variation (CV) %, which were as follows 15.3 (13.1-18.7), 5.5% and 47.7, (40.5-58.1), 6.4%, respectively. Linearity using patient samples and calibrators were measured from 201 to 16,177 and 188 to 30,000 pg/mL, respectively. The regression equation was y = -23.2 + 1.008x (SE = 162.4) for low levels and y = 779.9 + 1.006x (SE = 668) for high levels. Method comparison by Passing-Bablock analysis was as follows y = -209.77 + 1.047x (Syx  = 335.3). The correlation coefficient (95% CI) was 0.869 (0.772-0.927) with statistical significance (p < 0.001). Reference intervals from 120 normal healthy subjects showed that 300 pg/mL was the cut off. Presepsin tended to show a higher value at higher ages and in males. Presepsin showed correlation with some parameters, and the correlation coefficient (p value) were as follows hematocrit, 0.198 (0.03); eGFR (CKD-EPI), -0.240 (0.0129); MDRD-eGFR, -0.194 (0.048), respectively.

CONCLUSION:

Presepsin measurement by HISCL-5000 showed reliable performance. Further clinical studies are required for the diagnosis and prognosis of sepsis.