Transcatheter retrieval of atrial septal defect and patent ductus arteriosus occluder: a guidance for device retrieval based on comprehensive bench tests.
Cardiol Young
; 33(9): 1597-1605, 2023 Sep.
Article
em En
| MEDLINE
| ID: mdl-36093848
OBJECTIVES: The aim of this study is to establish a guidance for device retrieval based on comprehensive bench tests. BACKGROUND: Device embolisation remains a major complication in transcatheter closure of atrial septal defect and patent ductus arteriosus. Although percutaneous retrieval is feasible in the majority of cases, surgical retrieval may be required in complicated circumstances. However, the methods of transcatheter device retrieval have not been completely established. METHODS: Bench tests of device retrieval were performed to verify the appropriate retrieval method according to device type/size. The devices used for testing were Amplatzer Septal Occluder (Abbott, Chicago, IL, United States of America), Figulla Flex II (Occlutech GmbH, Jena, Germany), Amplatzer Duct Occluder-I (Abbott), Amplatzer Duct Occluder-II (Abbott), and Amplatzer Vascular Plug-II (Abbott). The retrieval equipment constituted diagnostic catheters (multipurpose catheter and right Judkins catheter, 4-Fr or 5-Fr, Gadelius Medical, Tokyo, Japan), delivery sheath and cables for each device, Amplatz goose neck snares (Medtronic, Minneapolis, MN, United States of America), OSYPKA CATCHER (Osypka ag, Rheinfelden-Herten, Germany), and OSYPKA LASSOS (Osypka). We investigated the retrieval equipment and sheath sizes required for a successful retrieval procedure for variously sized devices. RESULTS: For patent ductus arteriosus devices, the type of snare and the snaring position are considered important. For atrial septal defect devices, simple snare capture or a double-snare technique with a sufficiently large sheath is effective. Special care should be taken when using the OSYPKA CATCHER for device retrieval. CONCLUSIONS: The results of this study may assist in the selection of both capture devices and a retrieval sheath or a catheter for complete retrieval.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Permeabilidade do Canal Arterial
/
Dispositivo para Oclusão Septal
/
Comunicação Interatrial
Tipo de estudo:
Guideline
Limite:
Humans
Idioma:
En
Ano de publicação:
2023
Tipo de documento:
Article