Assessing component orientation of total hip arthroplasty using the low-dose bi-planar radiographs.
BMC Musculoskelet Disord
; 23(1): 886, 2022 Sep 26.
Article
em En
| MEDLINE
| ID: mdl-36154920
BACKGROUND: Three-dimensional computed tomography (3D CT) reconstruction is the reference standard for measuring component orientation. However, functional cup orientation in standing position is preferable compared with supine position. The low-dose bi-planar radiographs can be used to analyze standing cup component orientation. We aimed to assess the validity and reliability of the component orientation using the low-dose bi-planar radiographs compared with the 3D CT reconstruction, and explore the differences between the functional cup orientation in standing radiographs and supine CT scans. METHODS: A retrospective study, including 44 patients (50 hips) with total hip arthroplasty (THA), was conducted. CT scans were taken 1 week after surgery and the low-dose bi-planar radiographs were taken in the follow-up 6 weeks later. Component orientation measurement was performed using the anterior pelvic plane and the radiographic coronal plane as reference, respectively. RESULTS: The study showed no significant difference in cup anteversion (p = 0.160), cup inclination (p = 0.486), and stem anteversion (p = 0.219) measured by the low-dose bi-planar radiographs and 3D reconstruction. The differences calculated by the Bland-Altman analysis ranged from - 0.4° to 0.6° for the three measured angles. However, the mean absolute error was 4.76 ± 1.07° for functional anteversion (p = 0.035) and 4.02 ± 1.08° for functional inclination (p = 0.030) measured by the bi-planar radiographs and supine CT scans. CONCLUSIONS: The low-dose bi-planar radiographs are the same reliable and accurate as 3D CT reconstruction to assess post-THA patients' component orientation, while providing more valuable functional component orientation than supine CT scans.
Palavras-chave
Texto completo:
1
Base de dados:
MEDLINE
Assunto principal:
Artroplastia de Quadril
/
Prótese de Quadril
Tipo de estudo:
Observational_studies
Limite:
Humans
Idioma:
En
Ano de publicação:
2022
Tipo de documento:
Article