Your browser doesn't support javascript.
loading
High Incidence of Transient Permanent Pacemaker Rate After Rapid Deployment Valve Replacement: Insights of a 9-Year Single-Centre Experience.
Krasivskyi, Ihor; Djordjevic, Ilija; Ivanov, Borko; Gerfer, Stephen; Suhr, Laura; Avgeridou, Soi; Sabashnikov, Anton; Kuhn, Elmar; Rahmanian, Parwis; Mader, Navid; Eghbalzadeh, Kaveh; Wahlers, Thorsten.
Afiliação
  • Krasivskyi I; University Hospital Cologne, Heart Center Cologne, Department of Cardiothoracic Surgery, Cologne, Germany. Electronic address: ihor.krasivskyi@uk-koeln.de.
  • Djordjevic I; University Hospital Cologne, Heart Center Cologne, Department of Cardiothoracic Surgery, Cologne, Germany.
  • Ivanov B; University Hospital Cologne, Heart Center Cologne, Department of Cardiothoracic Surgery, Cologne, Germany.
  • Gerfer S; University Hospital Cologne, Heart Center Cologne, Department of Cardiothoracic Surgery, Cologne, Germany.
  • Suhr L; University Hospital Cologne, Heart Center Cologne, Department of Cardiothoracic Surgery, Cologne, Germany.
  • Avgeridou S; University Hospital Cologne, Heart Center Cologne, Department of Cardiothoracic Surgery, Cologne, Germany.
  • Sabashnikov A; University Hospital Cologne, Heart Center Cologne, Department of Cardiothoracic Surgery, Cologne, Germany.
  • Kuhn E; University Hospital Cologne, Heart Center Cologne, Department of Cardiothoracic Surgery, Cologne, Germany.
  • Rahmanian P; University Hospital Cologne, Heart Center Cologne, Department of Cardiothoracic Surgery, Cologne, Germany.
  • Mader N; University Hospital Cologne, Heart Center Cologne, Department of Cardiothoracic Surgery, Cologne, Germany.
  • Eghbalzadeh K; University Hospital Cologne, Heart Center Cologne, Department of Cardiothoracic Surgery, Cologne, Germany.
  • Wahlers T; University Hospital Cologne, Heart Center Cologne, Department of Cardiothoracic Surgery, Cologne, Germany.
Heart Lung Circ ; 31(12): 1658-1665, 2022 Dec.
Article em En | MEDLINE | ID: mdl-36155721
INTRODUCTION: The incidence of new permanent pacemaker implantation (PPI) after rapid deployment aortic valve replacement (RDAVR) remains debated. Expertise in this field has significantly increased over the last decade. This study aimed to investigate the need for PPI following implantation of a rapid deployment (RD) valve. METHODS: Analysis of n=372 patients who underwent Edwards INTUITY (Edwards Lifesciences, Irvine, CA, USA) (n=251) and Perceval (Sorin/LivaNova Group, Saluggia, Italy) (n=121) valve replacement at the current institution between May 2012 and August 2018 was performed. Coronary artery bypass graft procedures were additionally performed in patients with coronary artery disease. Baseline, preoperative and postoperative outcomes were examined regarding correctness and completeness of the procedure. Statistical analysis was performed using SPSS Version 23.0.0 (IBM Corp, Armonk, NY, USA). RESULTS: A total of 372 patients (aged 75±6.3 yrs) with a high grade of aortic valve stenosis underwent either Edwards INTUITY (67%) or Perceval (33%) valve replacement. Seventy-six (76) patients (20%) presented with preoperative conduction disorders. Sixty (60) patients (16%) underwent PPI, which in most cases was performed during the first month after the initial operation. Follow-up was performed up to 9 years, presenting a persistent pacemaker dependency rate of 50% among all patients who underwent PPI. Twenty (20) (40%) PPI recipients showed no dependency on pacemaker, while 10 (10%) required temporary pacemaker support. Mean length of ICU stay was 4±5.1 days and in-hospital stay was 8.2±7.6 days. CONCLUSIONS: The incidence of PPI after RD valve implantation still remains high compared with conventional aortic valve replacement. However, this study shows that this phenomenon appears to be transient in a significant proportion of the patients undergoing RD valve replacement. These findings might contribute to the scientific discussion and should be taken into consideration for the indication of RD valve replacement.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Estenose da Valva Aórtica / Marca-Passo Artificial / Próteses Valvulares Cardíacas / Implante de Prótese de Valva Cardíaca Tipo de estudo: Incidence_studies / Risk_factors_studies Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Estenose da Valva Aórtica / Marca-Passo Artificial / Próteses Valvulares Cardíacas / Implante de Prótese de Valva Cardíaca Tipo de estudo: Incidence_studies / Risk_factors_studies Limite: Humans Idioma: En Ano de publicação: 2022 Tipo de documento: Article