Small-field dosimetry with detector-specific output correction factor for single-isocenter stereotactic radiotherapy of single and multiple brain metastases.
Ono, Tomohiro; Kawata, Kohei; Nakamura, Mitsuhiro; Uto, Megumi; Mizowaki, Takashi.
Afiliação
  • Ono T; Department of Radiation Oncology and Image-Applied Therapy, Graduate School of Medicine, Kyoto University, 54 Kawahara-Cho, Shogoin, Sakyo-Ku, Kyoto, Kyoto, 606-8507, Japan. tono@kuhp.kyoto-u.ac.jp.
  • Kawata K; Division of Clinical Radiology Service, Kyoto University Hospital, 54 Kawahara-Cho, Shogoin, Sakyo-Ku, Kyoto, Kyoto, 606-8507, Japan.
  • Nakamura M; Division of Medical Physics, Department of Information Technology and Medical Engineering, Human Health Sciences, Graduate School of Medicine, Kyoto University, 53 Kawahara-Cho, Shogoin, Sakyo-Ku, Kyoto, Kyoto, 606-8507, Japan.
  • Uto M; Department of Radiation Oncology and Image-Applied Therapy, Graduate School of Medicine, Kyoto University, 54 Kawahara-Cho, Shogoin, Sakyo-Ku, Kyoto, Kyoto, 606-8507, Japan.
  • Mizowaki T; Department of Radiation Oncology and Image-Applied Therapy, Graduate School of Medicine, Kyoto University, 54 Kawahara-Cho, Shogoin, Sakyo-Ku, Kyoto, Kyoto, 606-8507, Japan.
Radiol Phys Technol ; 16(1): 10-19, 2023 Mar.
Article em En | MEDLINE | ID: mdl-36272022
ABSTRACT
Recently, the International Atomic Energy Agency and the American Association of Physicists in Medicine reported correction factors (CFs) for detector-response variation considering the uncertainty in detector readings in small-field dosimetry. In this study, the effect of CFs on small-field dosimetry measurements was evaluated for single-isocenter stereotactic radiotherapy for brain metastases. The output factors (OPFs) were measured with and without CFs in a water-equivalent sphere phantom using TrueBeam with a flattening-filter-free energy of 10 MV. Five detectors were used in a perpendicular orientation CC01, 3D pinpoint ionization chambers, unshielded SFD detector, shielded EDGE detector, and microDiamond detector. First, the square-field sizes were set to 5-100 mm using a multi-leaf collimator (MLC) field. The OPFs were evaluated in the presence and absence of CFs. Second, single-isocenter stereotactic irradiation was performed on 22 brain metastases in 15 patients following dynamic conformal arc (DCA) treatment. The equivalent field size was calculated using the MLC aperture for each planning target volume. For the OPFs, the mean deviations from the median of the doses measured with detectors other than the CC01 for square-field sizes larger than 10 mm were within ± 4.3% of the median without CFs, and ± 3.3% with CFs. For DCA plans, the deviations without and with CFs were - 2.3 ± 1.9% and - 4.8 ± 2.4% for CC01, - 1.1 ± 3.0% and 1.0 ± 1.6% for 3D pinpoint, 8.8 ± 3.0% and 2.9 ± 2.8% for SFD, - 3.1 ± 3.0% and - 13.5 ± 4.0% for EDGE, and 8.9 ± 2.1% and 0.8 ± 1.9% for microDiamond. This feasibility study confirmed that the deviation of the detectors can be reduced using an appropriate detector with CFs.