Technical features of vaping drug delivery system for bronchodilator delivery.
Chaoui, Mariam; Perinel-Ragey, Sophie; Prévôt, Nathalie; Leclerc, Lara; Pourchez, Jérémie.
Afiliação
  • Chaoui M; Mines Saint-Etienne, Univ Lyon, Univ Jean Monnet, Etablissement Français du Sang, INSERM, U 1059 Sainbiose, Centre CIS, F-42023 Saint-Etienne, France; INSERM, U1059, F-42023 Saint Etienne, France.
  • Perinel-Ragey S; INSERM, U1059, F-42023 Saint Etienne, France; Faculté de médecine, Université Jean Monnet, Saint Etienne, France; CHU Saint-Etienne, Medical-Surgical Intensive Care Unit, F-42055 Saint-Etienne, France.
  • Prévôt N; INSERM, U1059, F-42023 Saint Etienne, France; Faculté de médecine, Université Jean Monnet, Saint Etienne, France; CHU Saint-Etienne, Department of Nuclear Medicine, F-42055 Saint-Etienne, France.
  • Leclerc L; Mines Saint-Etienne, Univ Lyon, Univ Jean Monnet, Etablissement Français du Sang, INSERM, U 1059 Sainbiose, Centre CIS, F-42023 Saint-Etienne, France; INSERM, U1059, F-42023 Saint Etienne, France.
  • Pourchez J; Mines Saint-Etienne, Univ Lyon, Univ Jean Monnet, Etablissement Français du Sang, INSERM, U 1059 Sainbiose, Centre CIS, F-42023 Saint-Etienne, France; INSERM, U1059, F-42023 Saint Etienne, France. Electronic address: pourchez@emse.fr.
Int J Pharm ; 628: 122350, 2022 Nov 25.
Article em En | MEDLINE | ID: mdl-36374741
ABSTRACT

PURPOSE:

Several technical features influencing bronchodilator delivery were evaluated using different vaping drug delivery systems (VDDS).

METHODS:

Terbutaline in powder form, combined with 1, 3- propanediol used as e-liquid was tested at different concentrations (1 and 2.5 mg/mL), power levels (15 W and 30 W), and set applied resistances (0.15 to 1.5 O) to compare the efficiency of three VDDS (GS AIR2, GS TANK, CUBIS). Samples were collected with a Glass Twin Impinger (GTI). A High Performance Liquid Chromatography (HPLC) was used for drug quantification. The Next Generation Impactor (NGI) measured particle size distribution. Results were also considered with a clinical jet nebulizer (Cirrus TM 2, 2 mL of terbutaline at 2.5 mg/mL).

RESULTS:

GS AIR2 with resistance = 1.5 O; power = 15 W, and [Terbutaline] = 2.5 mg/mL represents the optimal VDDS conditions to deliver a respirable dose of 20.05 ± 4.2 µg/puff with a mass median aerodynamic diameter (MMAD) of 1.41 ± 0.03 µm. Thus, 52 puffs were required (lasting approximately 15 min of vaping) to reach similar respirable dose and MMAD compared to nebulization.

CONCLUSION:

We proved that several crucial VDDS technical parameters govern the performance of respiratory bronchodilator delivery including the resistance, power level and atomizer design.