Your browser doesn't support javascript.
loading
Development of a capillary electrophoretic method for determination of ketorolac enantiomers in human plasma using cationic ß-cyclodextrin derivative as a chiral selector.
Kravchenko, Anastasia V; Kolobova, Ekaterina A; Kechin, Arseniy A; Kartsova, Liudmila A.
Afiliação
  • Kravchenko AV; Institute of Chemistry, The Department of Organic Chemistry, Saint Petersburg State University, Saint Petersburg, Russia.
  • Kolobova EA; Institute of Chemistry, The Department of Organic Chemistry, Saint Petersburg State University, Saint Petersburg, Russia.
  • Kechin AA; The Federal State Institute of Public Health 'The Nikiforov Russian Center of Emergency and Radiation Medicine', The Ministry of Russian Federation for Civil Defence, Emergencies and Elimination of Consequences of Natural Disasters, Saint Petersburg, Russia.
  • Kartsova LA; Institute of Chemistry, The Department of Organic Chemistry, Saint Petersburg State University, Saint Petersburg, Russia.
J Sep Sci ; 46(2): e2200601, 2023 Jan.
Article em En | MEDLINE | ID: mdl-36420913
A novel approach for the separation of ketorolac enantiomers by capillary electrophoresis is presented. A cationic ß-cyclodextrin derivative based on imidazole was synthesized and used as a chiral selector in the background electrolyte. The influence of pH and ionic strength of background electrolyte, as well as cationic ß-cyclodextrin derivative concentration on the resolution of ketorolac enantiomers, was investigated. The highest value of the resolution for ketorolac enantiomers was 1.46 when the background electrolyte consisted of 25 mM NaH2 PO4 (pH 6.4) with 1 mM 1-butyl-3-ß-cyclodextrinimidazolium tosylate. Additionally, the possibilities of cationic derivatives for the separation of ketoprofen enantiomers were shown (peak resolution 1.06). The two-step preconcentration mode was developed to reduce the limit of detection of individual enantiomers. The proposed approach was successfully applied to determine ketorolac enantiomers in tablet "Ketorol express" and human plasma. The calibration range of ketorolac enantiomers for plasma samples was 0.25-2.50 µg/ml with coefficients of determination ≥ 0.99. The relative standard deviation both of the peak area and migration time was less than 15%, as well as the accuracy ranged from 90.1% to 110.2% for both analytes. The limits of detection were 44 and 55 ng/ml for R- and S-ketorolac. The quantity of ketorolac in plasma was verified with high-performance liquid chromatography.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Cetorolaco / Ciclodextrinas Limite: Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Cetorolaco / Ciclodextrinas Limite: Humans Idioma: En Ano de publicação: 2023 Tipo de documento: Article