Feasibility and safety of zero-fluoroscopy left bundle branch pacing: An initial experience.
Ramos-Maqueda, Javier; Melero-Polo, Jorge; Montilla-Padilla, Isabel; Ruiz-Arroyo, José Ramón; Cabrera-Ramos, Mercedes.
Afiliação
  • Ramos-Maqueda J; Arrhythmias Unit, Department of Cardiology, University Hospital Clinico Lozano Blesa, Zaragoza, Spain.
  • Melero-Polo J; Arrhytmias Unit, Instituto Investigación Sanitaria Aragón, Aragon, Spain.
  • Montilla-Padilla I; Arrhythmias Unit, Department of Cardiology, University Hospital Clinico Lozano Blesa, Zaragoza, Spain.
  • Ruiz-Arroyo JR; Arrhytmias Unit, Instituto Investigación Sanitaria Aragón, Aragon, Spain.
  • Cabrera-Ramos M; Arrhythmias Unit, Department of Cardiology, University Hospital Clinico Lozano Blesa, Zaragoza, Spain.
J Cardiovasc Electrophysiol ; 34(2): 429-436, 2023 02.
Article em En | MEDLINE | ID: mdl-36448425
ABSTRACT

INTRODUCTION:

Left bundle branch pacing (LBBP) has emerged in recent years as a new pacing modality, providing patients with a narrower paced QRS than conventional pacing and stable pacing parameters. At the same time, there is a growing concern about the use of fluoroscopy in pacemaker implantations, given its harmful effects on both patients and operators. However, there are no prior experiences of zero-fluoroscopy in LBBP procedure.

METHODS:

We conducted an observational prospective study recruiting consecutive patients that underwent zero-fluoroscopy LBBP pacemaker implantation. A 6-month follow-up visit was programmed for every patient. The main goal of our study was to assess the efficacy, feasibility, and safety of the procedure.

RESULTS:

From January 2021 to February 2022, we included 10 patients, 8 males. The average age was 63 ± 4 years. The procedure was successful in all patients. We observed a significant reduction in paced QRS width compared with basal QRS width (149 ± 31.9 vs. 116 ± 15.6 ms, p = .02). All device parameters remained stable at 6-month follow-up no significant differences in mean impedance (700.5 ± 136.4 vs. 494 ± 72.7 Ohm, p = .09), capture threshold (0.67 ± 0.2 vs. 0.83 ± 0.2 V @ 0.4 ms, p = .27) or endocardial V-wave amplitude (10.6 ± 5.2 vs. 13.9 ± 6.3 mV, p = .19). No complications were reported in any case.

CONCLUSION:

Zero-fluoroscopy LBBP is feasible and safe, and it may be considered in cases where radiation exposure is contraindicated or especially undesirable. Future randomized clinical trials are needed for the widespread use of this new technique.
Assuntos
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fascículo Atrioventricular / Estimulação Cardíaca Artificial Tipo de estudo: Clinical_trials / Observational_studies Limite: Aged / Humans / Male / Middle aged Idioma: En Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Base de dados: MEDLINE Assunto principal: Fascículo Atrioventricular / Estimulação Cardíaca Artificial Tipo de estudo: Clinical_trials / Observational_studies Limite: Aged / Humans / Male / Middle aged Idioma: En Ano de publicação: 2023 Tipo de documento: Article