Simultaneous Leadless Pacemaker and Subcutaneous ICD Implantation With Intraoperative Screening: Workflow in Two Patients.
Nieves, Jonathan; Laslett, David B; Basil, Anuj; Whitman, Isaac R; Cooper, Joshua M; Cronin, Edmond M.
Afiliação
  • Nieves J; Lewis Katz School of Medicine at Temple University, Philadelphia, Pennsylvania, USA.
  • Laslett DB; Lewis Katz School of Medicine at Temple University, Philadelphia, Pennsylvania, USA.
  • Basil A; Lewis Katz School of Medicine at Temple University, Philadelphia, Pennsylvania, USA.
  • Whitman IR; Lewis Katz School of Medicine at Temple University, Philadelphia, Pennsylvania, USA.
  • Cooper JM; Lewis Katz School of Medicine at Temple University, Philadelphia, Pennsylvania, USA.
  • Cronin EM; Lewis Katz School of Medicine at Temple University, Philadelphia, Pennsylvania, USA.
JACC Case Rep ; 4(23): 101535, 2022 Dec 07.
Article em En | MEDLINE | ID: mdl-36507298
A communicating subcutaneous implantable cardioverter-defibrillator (ICD) and leadless pacemaker system is being developed for patients who require both pacing and ICD therapy. It is important to ensure that the paced morphology from the leadless pacemaker will be sensed appropriately by the subcutaneous ICD. We present 2 cases illustrating our approach and workflow. (Level of Difficulty: Intermediate.).
Palavras-chave

Texto completo: 1 Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies / Screening_studies Idioma: En Ano de publicação: 2022 Tipo de documento: Article