Correcting image distortions from a nonlinear B 1 + $$ {\boldsymbol{B}}_{\mathbf{1}}
Wang, Paul; Froelich, Taylor; Torres, Efraín; DelaBarre, Lance; Jenkins, Parker; Radder, Jerahmie; Mullen, Michael; Garwood, Michael.
Afiliação
  • Wang P; Department of Biomedical Engineering, University of Minnesota, Minneapolis, Minnesota, USA.
  • Froelich T; Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota, Minneapolis, Minnesota, USA.
  • Torres E; Department of Biomedical Engineering, University of Minnesota, Minneapolis, Minnesota, USA.
  • DelaBarre L; Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota, Minneapolis, Minnesota, USA.
  • Jenkins P; Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota, Minneapolis, Minnesota, USA.
  • Radder J; Department of Biomedical Engineering, University of Minnesota, Minneapolis, Minnesota, USA.
  • Mullen M; Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota, Minneapolis, Minnesota, USA.
  • Garwood M; Center for Magnetic Resonance Research and Department of Radiology, University of Minnesota, Minneapolis, Minnesota, USA.
Magn Reson Med ; 89(5): 2100-2108, 2023 05.
Article em En | MEDLINE | ID: mdl-36517956
ABSTRACT

PURPOSE:

To correct image distortions that result from nonlinear spatial variation in the transmit RF field amplitude ( B 1 + $$ {B}_1^{+} $$ ) when performing spatial encoding with the method called frequency-modulated Rabi encoded echoes (FREE). THEORY AND

METHODS:

An algorithm developed to correct image distortion resulting from the use of nonlinear static field (B0 ) gradients in standard MRI is adapted herein to correct image distortion arising from a nonlinear B 1 + $$ {B}_1^{+} $$ -gradient field in FREE. From a B 1 + $$ {B}_1^{+} $$ -map, the algorithm performs linear interpolation and intensity scaling to correct the image. The quality of the distortion correction is evaluated in 1.5T images of a grid phantom and human occipital lobe.

RESULTS:

An expanded theoretical description of FREE revealed the symmetry between this B 1 + $$ {B}_1^{+} $$ -gradient field spatial-encoding and standard B0 -gradient field spatial-encoding. The adapted distortion-correction algorithm substantially reduced image distortions arising in the spatial dimension that was encoded by the nonlinear B 1 + $$ {B}_1^{+} $$ gradient of a circular surface coil.

CONCLUSION:

Image processing based on straightforward linear interpolation and intensity scaling, as previously applied in conventional MRI, can effectively reduce distortions in FREE images acquired with nonlinear B 1 + $$ {B}_1^{+} $$ -gradient fields.