An artificial intelligence-driven image quality assessment system for whole-body [18F]FDG PET/CT.
Qi, Chi; Wang, Shuo; Yu, Haojun; Zhang, Yiqiu; Hu, Pengcheng; Tan, Hui; Shi, Yonghong; Shi, Hongcheng.
Afiliação
  • Qi C; Department of Nuclear Medicine, Zhongshan Hospital, Fudan University, Shanghai, China.
  • Wang S; Institute of Nuclear Medicine, Fudan University, No. 180 in Fenglin Road, Shanghai, China.
  • Yu H; Shanghai Institute of Medical Imaging, Shanghai, China.
  • Zhang Y; Cancer Prevention and Treatment Center, Zhongshan Hospital, Fudan University, Shanghai, China.
  • Hu P; Digital Medical Research Center of School of Basic Medical Sciences, Fudan University, 138 Yixueyuan Road, Shanghai, China.
  • Tan H; Shanghai Key Laboratory of Medical Image Computing and Computer Assisted Intervention, Shanghai, China.
  • Shi Y; Department of Nuclear Medicine, Zhongshan Hospital, Fudan University, Shanghai, China.
  • Shi H; Institute of Nuclear Medicine, Fudan University, No. 180 in Fenglin Road, Shanghai, China.
Eur J Nucl Med Mol Imaging ; 50(5): 1318-1328, 2023 04.
Article em En | MEDLINE | ID: mdl-36529840
ABSTRACT

PURPOSE:

Image quality control is a prerequisite for applying PET/CT. This study aimed to develop an artificial intelligence-driven real-time and accurate whole-body [18F]FDG PET/CT image quality assessment system.

METHODS:

This study included 173 patients (age, 59 ± 12 years; 66.3% males) with whole-body [18F]FDG PET/CT imaging. Images of ten patients were used as an educational set. Images of the rest 163 patients were reconstructed to 952 images by simulating several scanning times and randomly split into training (60%, 98 patients, 578 images), validation (20%, 33 patients, 192 images), and test (20%, 32 patients,182 images) sets. Two experienced physicians (R1 and R2) independently assessed the image quality of thorax, abdomen, and pelvis region twice (R1a and b; R2a and b), 1 month apart, using a 5-point Likert scale. Objective image quality metrics were extracted from the mediastinal blood pool, three liver levels, and the bilateral gluteus maximus. The developed convolutional neural networks for image quality assessment (IQA-CNNs) generated the subjective quality scores and objective image metrics. The IQA-CNNs and physicians' performances were compared for localization accuracy, score agreement, and process time.

RESULTS:

The physicians demonstrated good inter- and intra-rater subjective assessment agreement, with kappa coefficients (R1a vs. R2a, R1a vs. R1b, R2a vs. R2b, and R1a vs. R2b) of 0.78, 0.77, 0.76, and 0.80. The IQA-CNNs and R1 or R2 agreed in the subjective assessments, with kappa coefficients of 0.79 and 0.78. IQA-CNNs and R1 or R2 also agreed in their objective image quality assessment (ICC > 0.60). The IQA-CNNs evaluation speed was 200 times faster than the manual assessment.

CONCLUSION:

An automated system for rapid assessment of [18F]FDG PET/CT image quality was developed, showing comparable performance to senior physicians. The system generates a comprehensive and detailed image quality assessment report, including subjective visual scores and objective image metrics for various anatomical regions.