Analytical and clinical performances of seven direct detection assays for SARS-CoV-2.
Matsumura, Yasufumi; Yamazaki, Wataru; Noguchi, Taro; Yamamoto, Masaki; Nagao, Miki.
Afiliação
  • Matsumura Y; Department of Clinical Laboratory Medicine, Kyoto University Graduate School of Medicine, 54 Shogoin-kawaharacho, Sakyo-ku, Kyoto, 6068507, Japan.
  • Yamazaki W; Center for Southeast Asian Studies, Kyoto University, 46 Yoshida Shimoadachi-cho, Sakyo-ku, Kyoto, 6068501, Japan.
  • Noguchi T; Department of Clinical Laboratory Medicine, Kyoto University Graduate School of Medicine, 54 Shogoin-kawaharacho, Sakyo-ku, Kyoto, 6068507, Japan.
  • Yamamoto M; Department of Clinical Laboratory Medicine, Kyoto University Graduate School of Medicine, 54 Shogoin-kawaharacho, Sakyo-ku, Kyoto, 6068507, Japan.
  • Nagao M; Department of Clinical Laboratory Medicine, Kyoto University Graduate School of Medicine, 54 Shogoin-kawaharacho, Sakyo-ku, Kyoto, 6068507, Japan.
J Clin Virol Plus ; 3(1): 100138, 2023 Feb.
Article em En | MEDLINE | ID: mdl-36683610
ABSTRACT

Background:

Direct detection tests for severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) that bypass complicated nucleic acid/antigen purification steps are promising tools for the rapid diagnosis of coronavirus disease 2019 (COVID-19).

Methods:

To determine the analytical and clinical diagnostic performances of the direct detection assays, we compared 6 direct molecular detection assays, including two loop-mediated isothermal amplification (LAMP) assays and one lateral flow antigen assay, against the reference extraction-based RT-PCR assay using 183 respiratory samples (87 nasopharyngeal swabs, 51 saliva samples, and 45 sputum samples).

Results:

Analytical sensitivity analysis showed that the direct RT-PCR assay of Toyobo exhibited the lowest LOD of 1,000 copies/mL. Compared with the 80 positive and 103 negative samples based on the reference assay, the Toyobo assay had the highest positive percent agreement (PPA) of 96.3%, followed by the two direct RT-PCR assays of Takara and Shimadzu and one LAMP assay of Eiken (86.3-87.5%). The Fujirebio antigen assay had the lowest PPA of 44.7% among the assays tested. The negative percent agreement of these direct detection assays was 100%, except for the Eiken assay (96.3%).

Conclusions:

Large differences in PPA existed among the direct detection tests. Laboratories need to take these characteristics into consideration before implementing these assays.