Calibration of the Chemcatcher® passive sampler and derivation of generic sampling rates for a broad application in monitoring of surface waters.
Römerscheid, Mara; Paschke, Albrecht; Schneider, Selma; Blaha, Maximilian; Harzdorf, Julia; Schüürmann, Gerrit.
Afiliação
  • Römerscheid M; UFZ Department of Ecological Chemistry, Helmholtz Centre for Environmental Research, Permoserstr. 15, 04318 Leipzig, Germany; Institute of Organic Chemistry, Technical University Bergakademie Freiberg, Leipziger Str. 29, 09599 Freiberg, Germany. Electronic address: mara.roemerscheid@ufz.de.
  • Paschke A; UFZ Department of Ecological Chemistry, Helmholtz Centre for Environmental Research, Permoserstr. 15, 04318 Leipzig, Germany.
  • Schneider S; UFZ Department of Ecological Chemistry, Helmholtz Centre for Environmental Research, Permoserstr. 15, 04318 Leipzig, Germany.
  • Blaha M; UFZ Department of Ecological Chemistry, Helmholtz Centre for Environmental Research, Permoserstr. 15, 04318 Leipzig, Germany.
  • Harzdorf J; UFZ Department of Ecological Chemistry, Helmholtz Centre for Environmental Research, Permoserstr. 15, 04318 Leipzig, Germany.
  • Schüürmann G; Institute of Organic Chemistry, Technical University Bergakademie Freiberg, Leipziger Str. 29, 09599 Freiberg, Germany.
Sci Total Environ ; 871: 161936, 2023 May 01.
Article em En | MEDLINE | ID: mdl-36746283
ABSTRACT
We determined sampling rates for 34 pesticides, five pesticide transformation products, and 34 pharmaceutical compounds with the Chemcatcher (CC) passive sampler in a laboratory-based continuous-flow system at 40 cm/s and ambient temperature. Three different sampling phases were used styrene divinylbenzene disks (SDB-XC), styrene divinylbenzene reversed phase sulfonate disks (SDB-RPS), and hydrophilic lipophilic balance disks (HLB), in all cases covered with a diffusion-limiting polyethersulfone membrane. The measured sampling rates range from 0.007 L/d to 0.193 L/d for CC with SDB-XC (CC-XC), from 0.055 L/d to 0.796 L/d for CC with SDB-RPS (CC-RPS), and from 0.018 L/d to 0.073 L/d for CC equipped with HLB (CC-HLB). Comparison with sampling rates from literature enabled to derive generic sampling rates that can be used for compounds with unknown uptake kinetics such as transformations products and new compounds of interest. Field trial results demonstrate that the presently derived generic sampling rates are suitable for estimating time-weighted average concentrations within reasonable uncertainty limits. In this way, Chemcatcher passive sampling can be applied approximately to a broad range of solutes without the need for deriving compound-specific sampling rates, which enable compliance checks against environmental quality standards and further risk assessment.